Min. Order: | 1 Set/Sets |
---|---|
Trade Term: | FOB |
Payment Terms: | T/T |
Supply Ability: | 500Sets/per Month |
Place of Origin: | Chongqing |
Location: | Chongqing, China (Mainland) |
---|---|
Business Type: | Manufacturer, Trading Company |
Model No.: | GD-261A |
---|---|
Means of Transport: | Ocean, Air, Land |
Flash Point Tester: | Open-Cup Flash Point Tester |
Pensky-Martens Tester: | Closed-Cup Flash Point Tester |
Closed-Cup Flash Point Tester: | Pensky-Martens Closed-Cup Flash Point Tester |
Production Capacity: | 500Sets/per Month |
Packing: | Wooden Case |
Delivery Date: | 10days |
Introduction:
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.
Features:
l Light-touch colored LCD. English man-machine dialog interface. All parameters including preset flash temperature, sample number, atmospheric pressure and date can be set by prompt menu.
l Simulation tracking of temperature rising curve with time. Prompt function. Test date and time can be modified.
l 120 groups of data can be saved.
l Automatically calculate the corrected flash point when the pressure differs from standard atmospheric pressure.
l Differential detection. Automatic correct the system bias.
l Automatically open cover, ignite, detect and print result.
l Safe and easy to operate.
Specifications:
Power supply | AC(220±10%)V, 50Hz |
Flash point determination | Range: Ambient to 300℃ Meet requirement of GB/T 261-2008 Meet requirement of GB/T 261-2008 Resulotion: 0.1℃ |
Heating rate | Pr℃edure A: (5~6)℃/min Pr℃edure B:(1~1.5)℃/min Automatic control and manually adjustable |
Stirring rate | Pr℃edure A: (90~120)RPM Pr℃edure B: (250±10)RPM Automatic control and manually adjustable |
Igniting mode | Igniting source: gas (or other civilian fuels) Electric ignition. Gas flame diameter:3.2mm~4.8mm |
Ambient temperature | ≤ 40℃ |
Relative humidity | ≤ 80% |
Maximum power consumption | 500W |
Overall dimension | 410*360*310mm |