Automatic Pensky-Martens Closed Cup Flash Point Tester

Min. Order: 1 Set/Sets
Trade Term: FOB
Payment Terms: L/C, T/T, WU
Supply Ability: 1000/month
Place of Origin: Chongqing

Company Profile

Location: Chongqing, China (Mainland)
Business Type: Manufacturer, Trading Company
Main Products: Oil Purifier, Oil Tester

Product Detail

Model No.: GD-261D
Means of Transport: Ocean, Air, Land
Usage: Others
Power: Electronic
Brand Name: gold
Production Capacity: 1000/month
Packing: Veneer case
Delivery Date: within 15 days
Show

Product Description

I.Test Method

  • It conforms to ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester

  • For flash point determinations of fuels, lubricating oils, liquids containing suspended solids and liquids that tend to form a surface film during testing.

  • II.Features and Benefits

  • Conforms to ASTM D93 and related specifications

  •  Simple automation routine for easy operation

  • Micro-cumputer controlled system

  • Gas or electric ignition

  •  Flash point operation range between 0 and 300°C

  • Large viewing screen for observing test status at a distance from the unit

  • Automatic test process

  • with RS232 interface

  • Automatic barometric correction

  • Power supply : AC 220V±10%, 50Hz (can be customized)

  • III.Automatic Pensky-Martens Closed Cup Flash Point Tester

  • The automated Pensky-Martens flash point tester accurately determines the lowest flash point temperature of fuels, lubricating oils, and homogenous liquids (ASTM D93 A), or liquids containing suspended solids as well as liquids that tend to form a surface film during testing (ASTM D93 B).

  • Flash point tests are simply conducted by mounting the flash cup filled with sample into the test position and selecting a pre- or user-programmed test method.

  • A quick search method allows for determination of flash points for unknown samples and a method for asphalts is also included.

  • The automation routines provide accurate test results, even with users inexperienced in flash point test methods.

  • The flash point test result is automatically corrected to standard pressure (101.3 kPa).

  • The unit is equipped with a differential Pt-100 RTD probe designed to duplicate the response time of a mercury-in-glass thermometer as per ASTM D93-02a and E1-03a.

  • The system features multiple sensors for continually monitoring of instrument function and displaying an error message if a problem is detected.

  • The performance of the electrical ignitor is continuously checked, and the user is notified upon the need of replacement due to either damage or the end of its useful life.

  • The system is easily interfaced with an external PC for operation and method updates.

  • When performing a test, the system will display the stirring speed, temperature curve (also printed out), and current test status.

  • The system alerts the user is also alerted if the first application of the ignitor results in a flash or if no flash point is detected at the end of the test program.

  • If a flash is not detected 30°C above the expected flash point or at 300°C, then the test is automatically aborted for safety.

  •  An easy connection to the air ventilation system or external water connection provides a quick cool down between test runs for operational efficiency.

  • IV. Dimensions and Weight

    69*52*44cm,27kg 

     

    V.Same series products:

    ItemGD-261GD-261BGD-261AGD-261DGD-5208
    Testing Rangeambient~300°C  ambient~250°C  ambient~250°C  ambient~300°C  -30~50°C
    Repeatability≤2 °C≤2 °C   ≤2 °C  ≤2 °C≤2 OR 3°C






    Automatic up/down No No YesYes No
    SampleStandard oil cupStandard oil cupStandard oil cupStandard oil cup2ml
    With micropriter No Yes Yes Yes Yes 
    Communication port No No No YesNo
    Reproductility ≤4 °C  ≤4 °C  ≤4 °C  ≤4 °C  ≤2 OR 3 °C  
    Screen Display No No Yes Yes Yes 

GD-261D Automatic Pensky-Martens Closed Cup Flash Point Tester 