user login

Automatic Pensky-Martens Closed-Cup Flash Point Tester

2.0 USD
Min. Order: 1 Set/Sets
Trade Term: FOB
Payment Terms: Paypal, D/P, T/T, Money Gram
Supply Ability: 500Sets/per Month
Place of Origin: Chongqing

Company Profile

Location: Chongqing, China (Mainland)
Business Type: Trading Company
Main Products: Electrical Tester/Petroleum Products

Product Detail

Model No.: GD-261A
Means of Transport: Ocean, Air, Land
Standard: GB/T261,ASTM D93
Test sample: Gasoline,jet fuel,kerosene,diesel,gas turbine
Temperature range: Ambient~250
ignition mode: Manual
Temperature display: Color LCD
Printing: NO
Communication port: NO
Sample cup: Standard oil cup
Dimension: 410*360*310MM
Production Capacity: 500Sets/per Month
Packing: Wooden Case
Delivery Date: 10days
Show

Product Description

GD-261A Automatic Pensky-Martens Closed-Cup Flash Point Tester

Introduction:

This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.

 

Features:

l Light-touch colored LCD. English man-machine dialog interface. All parameters including preset flash temperature, sample number, atmospheric pressure and date can be set by prompt menu.

l Simulation tracking of temperature rising curve with time. Prompt function. Test date and time can be modified. 

l 120 groups of data can be saved.

l Automatically calculate the corrected flash point when the pressure differs from standard atmospheric pressure. 

l Differential detection. Automatic correct the system bias. 

l Automatically open cover, ignite, detect and print result. 

l Safe and easy to operate. 

 

Specifications:

Power supply

AC(220±10%)V,  50Hz

Flash point determination

Range: Ambient to 300℃

Meet requirement of GB/T 261-2008

Meet requirement of GB/T 261-2008

Resulotion: 0.1℃

Heating rate

Pr℃edure A: (5~6)℃/min

Pr℃edure B:(1~1.5)℃/min

Automatic control and manually adjustable

Stirring rate

Pr℃edure A: (90~120)RPM

Pr℃edure B: (250±10)RPM

Automatic control and manually adjustable

Igniting mode

Igniting source: gas (or other civilian fuels)

Electric ignition. Gas flame diameter:3.2mm~4.8mm

Ambient temperature

≤ 40℃

Relative humidity

≤ 80%

Maximum power consumption

500W

Overall dimension

 410*360*310mm

 


Post Buying Request