Introduction:
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.
Features:
l Light-touch colored LCD. English man-machine dialog interface. All parameters including preset flash temperature, sample number, atmospheric pressure and date can be set by prompt menu.
l Simulation tracking of temperature rising curve with time. Prompt function. Test date and time can be modified.
l 120 groups of data can be saved.
l Automatically calculate the corrected flash point when the pressure differs from standard atmospheric pressure.
l Differential detection. Automatic correct the system bias.
l Automatically open cover, ignite, detect and print result.
l Safe and easy to operate.
Specifications:
Power supply | AC(220±10%)V, 50Hz |
Flash point determination | Range: Ambient to 300℃ Meet requirement of GB/T 261-2008 Meet requirement of GB/T 261-2008 Resulotion: 0.1℃ |
Heating rate | Pr℃edure A: (5~6)℃/min Pr℃edure B:(1~1.5)℃/min Automatic control and manually adjustable |
Stirring rate | Pr℃edure A: (90~120)RPM Pr℃edure B: (250±10)RPM Automatic control and manually adjustable |
Igniting mode | Igniting source: gas (or other civilian fuels) Electric ignition. Gas flame diameter:3.2mm~4.8mm |
Ambient temperature | ≤ 40℃ |
Relative humidity | ≤ 80% |
Maximum power consumption | 500W |