Multi-Parameter Patient Monitor

Min. Order: 1 Set/Sets
Trade Term: FOB
Payment Terms: L/C, T/T, WU
Supply Ability: 1000 sets per month
Place of Origin: Beijing

Company Profile

Location: Beijing, China (Mainland)
Business Type: Manufacturer

Product Detail

Model No.: CD2000
Means of Transport: Air
Color: White
Size: 327*316*215mm
Standard configuration: ECG, SpO2, HR, RESP, TEMP,NIBP
Optional: EtCO2,IBP, Thermal printer,Net work connection
Display: 10.4"
Service: OEM/ODM
Production Capacity: 1000 sets per month
Packing: one device in each carton
Delivery Date: one week
Show

Product Description

Multi-Parameters Patient Monitors CD2000
Display: 10.4"
Dimension:327*316*215 mm
Standard Configuration: ECG, SpO2, HR, RESP, TEMP, NIBP
Optional: EtCO2, IBP, Thermal printer, Network connection
Features
Suitable for use in ICU/CCU
10.4"Big font display
Filled waveform display

Technical Specification
ECG/HR
Lead Type: 3-5 lead
Lead Selection:
ECG Waveform: 1-7 channel
Gain Selection: ×0.25, ×0.5, ×2, auto
Sweep Speed: 12.5mm/s, 25mm/s, 50mm/s
Heart Rate Range: Adult15~300 BPM,
Pediatric/Neonatal: 15~300 BPM
Accuracy: 1BPM or ±1%, whichever is greater
Resolution: 1 BPM
Protection: Withstand 4000VAC/50Hz voltage isolation
CMRR: Diagnostic mode ≥90dB
Respiration
Method: RA-LL impedance
Measurement Range: Adult: 0~120 Brpm
                  Pediatric/Neonatal: 0~150BrPM
Resolution: 1BrPM
Accuracy: ±2% or ±2BrPM, whichever is greater
Alarm Range: Adult: 0~120BrPM
Pediatric/Neonatal: 0~150BrPM
NIBP
Measurement unit: mmGh/kPa
Measurement Range: Adult10~270 mmHg
                 Pediatric: 10~200mmHg
                 Neonatal: 10~135mmHg
Resolution: 1mmHg
Accuracy: Maximum Mean error: 5mmHg
Maximum Standard deviation: 8mmHg
Temperature
Measurement Range:0~50℃ or 32~122℉
Resolution: 0.1℃ or 0.1℉
Accuracy: ±0.1℃ (0℃to50℉), exclusive of probe errors
Channel: 2 channel
SpO2
Range: 0~100%
Accuracy: 70%~100%; ±2%
Pulse Rate
Range: 20~300 BPM
Accuracy: ±2BPM or 2%, whichever is greater