Product Description
Main specification:
*Working condition
Power:AC 220V±10%; 50Hz±2%
Intake power rate:30W
Gas source:O2, N2O
rated working pressure:0.4±0.1MPa
*Flowmeter adjust range
O2:0.1~10L/min;N2O:0.1~10L/min
*Flush oxygen:>35L/min,<75L/min
*Enflurane(Isoflurane)Vaporizer:concentration:0~5Vol%
*Ventilator parameter
Tidal volume adjust range:0.1~1.00L
Frequency adjust range:10~30BPm
I:E ratro:1:1.5,1:2,1:3
Airway peak pressure:1~6KPa
*Alarm
Tidal volume:no tidal volume after 6 second,audible and visible alarm
*Airway peak pressure:
Up limited setting range:2~6KPa
Down limited setting:0.4KPa
When the airway peak pressure displaying & #118alue is over the limited setting & # 118alue or less than 0.4KPa,the visible and audible alarm and brightly in screen.
*The safety alarm for lacking of oxygen
When O2 pressure is less than 0.2MPa,the audible and visible alarm. If the time is going on for 7 seconds.N2O is cut off.