Product Description
Main Feature:
Be controlled by mircoprocess
Used for adult and child
IPPV and PVC for ventilation
End-inspiratory hold function
Positive end-expiratory pressure ventilation function
I:E ventilation
Compliance compensated function.The ventilator will compensated for tidal volume automatically when the airway pressure and patient lung compliance have been changed and the pipenline is leakage.
Back-up battery will be last for at least two hours.
Advance automatical control technology. high accuracy imported sensor,imported sensor,imported control valve,the bellow don't need to be changed if for child.
GMS absorber; good hermatically,easy using ,easy cleanly and sterilizelly.
Main specification:
Gas flow: O2 0.1-1L/min 1.25-10L/min
N20 0.1-1L/min 1.25-10L/min
Air 0.1-10/Lmin
O2,N2O chain link,O2 concentration is not less then 25% when using N2O
Flush O2:35-75L/min
Vaporizer(Enflurane,lsoflurane)
Temperature,flow and pressure are compensated Adjust range:0-VOL5%
Ventilation mode:
--IPPV --PVC
--SIMV --Manual
Ventilation settings:
Frequency:6-60/min I/E ratio:1:3-2:1
Tidal volume:50-1200ml Control pressure;1-6kPa
EIP:0-50% Trigger sensitive:-2.0- +2.0kPa
Ventilation parameter:
LED display:tidal, minute volume,frequency,airway pressure ,O2 concentration
Display picture:time-airway pressure wave picture manual ventilation tidal volume picture.
Safe alarm:
Oxygen is not full,ocer airway pressure limited,over minute volume limitation,over O2 concentration,no tidal volume,power shut off ,back-up battery is wrong