user login

Photoacoustic Spectrum Gas Analyzer in Oil

1.0 USD
Min. Order: 1 Set/Sets
Payment Terms: Paypal, L/C, T/T, WU
Supply Ability: 100 sets per yaer
Place of Origin: Beijing

Company Profile

Location: Beijing, China (Mainland)
Business Type: Manufacturer, Trading Company

Product Detail

Model No.: DK-200F
Means of Transport: Air
Brand Name: DukeTech
Production Capacity: 100 sets per yaer
Packing: wooden box
Delivery Date: 60 working days
Show

Product Description

Technology:

The DK-200F oil gas analyzer can realize gas analysis in multi-component oil, and the high-sensitivity measurement of short-path optical path can be realized by Finnish interference company's photoacoustic spectroscopy technology. Ultra-high sensitivity based enhanced cantilever beam Optical microphone technology enables linear dynamic range ratios up to 5 steps. Ultra-high sensitivity is achieved with 7 optical narrow-band filters. Extremely narrow optical filtering removes background gas interference; multiple spectral regions can be beneficial The smallest cross-sensitivity. The multi-gas analyzer is based on infrared photoacoustic spectroscopy technology. It has a pulsed infrared light source and passes through a narrow-band optical filter to form seven spectral bands in the mid-infrared region for gas analysis.


Characteristic:

  • Sub-ppm detection limit

  • The oil and gas separation speed is fast, and special environmental adaptation technology is adopted to eliminate the influence of temperature and humidity changes on the gas distribution coefficient.

  • High anti-interference performance, electromagnetic compatibility performance meets gb/t17626 and iec61000 standards

  • Provide two levels of alarm function, alarm signal can be transmitted remotely

  • Open database for access to power system LAN

  • Faster analysis cycle with a minimum monitoring period of 10 minutes, which can be set by the user. The recommended detection period is 12 hours.

  • The system has a secondary gas threshold alarm and a growth rate alarm function, and the alarm concentration and growth rate can be set arbitrarily.

  • Equipped with an expert system for judging latent faults, the latest fault diagnosis technology uses the iec60599 standard and the improved fuzzy three-ratio method, David triangle, cubic graphic method and other methods.

  • It has historical data trend graph query, report format display and printing function, which is convenient for longitudinal comparison of monitoring data to analyze transformer operation and fault development trend.

  • Support iec 61850 communication standard, optical fiber or twisted pair optional

  • High degree of system integration, optional with core ground current and ambient temperature and humidity monitoring function

  • Double-circuit multi-mode constant temperature control, temperature control accuracy of ± 0.1℃, equipment equipped with automatic constant temperature industrial air conditioner

  • Functional interface circuit adopts optocoupler isolation design to further improve system anti-interference performance

  • Full digital, remote data transmission, control and parameter settings with Ethernet

  • Highly scalable for easy integration with other monitoring devices

  • Analysis software man-machine interface friendly, full Chinese interface, easy to operate, rich system features

  • The core unit is a standard 19-inch rack structure with a compact structure, small size, easy maintenance and group screen

 

Technical parameter:

Technical specifications


Range


Gas

Lower limit / ppm

Upper limit / ppm


Hydrogen

5

5,000


Methane

1

50,000


Ethane

1

50,000


Ethylene

1

50,000


Acetylene

1

50,000


Carbon monoxide

5

50,000


Carbon dioxide

5

50,000


Oxygen

0%

100%


Nitrogen

0%

100%


Moisture measurement with micro water

0%

100%


Physical and mechanical


Ambient operating temperature

+5 - +50℃


Operating sample temperature

0 - +45℃


Storage temperature

-30°C - +80℃ 


Humidity

90% (non-condensing @30℃)


Vibration tolerance

IEC 60068-2-6, Fc test


Drop/Shock

Iec 60068-2-31 and ista 1a


Main power input frequency range

47-63Hz


Main power input voltage range

Universal 90-250vac


Weight

< 15kg


Size

1210pxW x 347.5pxH x 1100pxD (19.1In W x 5.5 in H x 17.3 in D)


Gas chamber volume

< 30ml


Oil volume in the extraction unit

<= 50ml


Dynamic range

500-50,000


Zero drift

± detection limit / 1 month


Measurement repeatability

5% of the measured value under reference conditions


Range drift

5% of measured value / 12 months


Precision

+/- 5% or +/- 2 ppm (whichever is greater)

+/-1% rh (H2O relative humidity is 0-90%)


 

standard

EN61326-1:2006

Radiation emission (A level )

EN61326-1:2006

Conducted emission (A level)

EN61000-3-2:2000

Steady state and wave harmonics

EN61000-3-3:2001

Flicker test

IEC61000-4-2:2001

Electrostatic discharge immunity test

IEC61000-4-3:2002

Radiation immunity

IEC61000-4-4:2001

Fast electrical transient/burst immunity test

IEC61000-4-5:2001

Surge immunity

IEC61000-4-6:2001

Conducted radio frequency immunity

IEC61000-4-8:2001

Magnetic field immunity

IEC61000-4-11:2001

Voltage dips and interruptions

Other

EN 61010-1:2001

IEC61010-1

UL61010-1

CAN/CSA-C22.2 No.61010-1

EN 50270:2006 


Post Buying Request