100V 50 - 60Hz Intense Pulsed Light Laser Hair Removal Equipment with Double Handles

Place of Origin: Beijing

Company Profile

Location: Beijing, China (Mainland)
Business Type: Manufacturer

Product Detail

Model No.: MED-220

Product Description

100V 50 - 60Hz Intense Pulsed Light Laser Hair Removal Equipment with Double Handles

 


Description:

100V 50 - 60Hz Intense Pulsed Light Laser Hair Removal Equipment with 250W, Double Handles MED-220

 

Applications

· 750-1200nm:   Hair Removal (Skin Type  IV)

· 690-1200nm:   Hair Removal (Skin Type  III)

· 640-1200nm:   Hair Removal (Skin Type  I & II ) ; Wrinkle Removal

· 590-1200nm:   Pigmentation and Vascularity; Skin Rejuvenation

· 560-1200nm:   Vascular and Pigmented Lesions; Skin Rejuvenation

· 530-1200nm:   Skin Rejuvenation; Superficial Vascular Lesions

· 510-1200nm:   Superficial Vascular Lesions

· 480-1200nm:   Light Pigmentation

· 430-1200nm:   Acne

· RF Treatment:  Skin Tightening and Face Lifting


 

Handpieces:





How it works?

 

E-Light is the perfect combination of MED-IPL technology with Bi-polar Radio Frequency (RF) and Skin Contact Cooling.

Radio Frequency is independent of melanin and thus independent of skin type. With Radio Frequency, increased selectivity and deeper dermal penetration would be achieved, thus lower energy is applied during the IPL treatment. Uncomfortable feeling during the IPL treatment will be decreased significantly and better result can be expected.  That is, less optical energy (IPL), when synergistically combined with Bi-polar RF, improves the safety factor when treating all skin types.

Meanwhile, contacting surface cooling technology applied on the treatment probe head can eliminate the heat effect attributed to optical energy, and increase the resistance of surface skin, reduce the RF absorption of the surface skin, greatly improving both the treatment performance and safety.  The technology is equally applicable and effective to mild color target tissues.

Laser technology has greatly improved the ability to treat melanocytic lesions and tattoos with the rapidly pulsed Q-switch neodymium: yttriumaluminiumgarnet (Nd: YAG) laser. The laser treatment of pigmented lesions and tattoos are based on principle of selected photothermolysis. The QS laser Systems can successfully lighten or eradicate a variety of benign epidermal and dermal pigmented lesions and tattoos with minimal risk of untoward effects.

 

 




 

Unique Advantages


2 in 1 device with double handles for different treatment items Magnative values make water-electric isolated, stable working Monopolar and Bipolar RF for your option.
 



 

Technical Specifications:

Post Buying Request

Wavelength (Spectrum)

530/640-1200nm (standard)430/480/560/590/690/750-1200nm (optional)

Energy Density (Fluence)

10-60J/cm2

RF Energy

Up to 50W

Spot Size

12x33mm² + 15x50mm² (standard)15x35mm² (optional)

Pulse Repetition Rate

0.2Hz; 2Hz

Pulse Duration

0.5-15ms

Pulse Delay

1-50ms

Pulses

1-15 pulses

IPL Peak Power

2000W

RF Power

Max. 100W 

Cooling