Min. Order: | 1 Set/Sets |
---|---|
Payment Terms: | Paypal, L/C, D/P, D/A, T/T, WU, Money Gram |
Supply Ability: | 100 Set/Sets per Month |
Place of Origin: | Chongqing |
Location: | Chongqing, China (Mainland) |
---|---|
Business Type: | Manufacturer, Trading Company, Other |
Model No.: | GD-259 |
---|---|
Means of Transport: | Ocean, Air, Land |
Usage: | Auto Testing Machine |
Power: | Electronic |
Brand Name: | GOLD |
Usage: | Water Soluble Acid Analyzer |
Specimen: | Petroleum products,additive,lubricating grease,wax and waxy products. |
Power supply: | AC 220 V±10%, 50 Hz |
Maximum power consumption: | 1200W |
Full scale: | 0~14.00 pH |
Standard: | GB/T 259 ,ASTM D974 |
Condition: | New |
Warranty: | 1 Year |
Function: | Water Soluble Acid Number amd Base Number Tester |
Production Capacity: | 100 Set/Sets per Month |
Packing: | Standard export carton, as per your requirements |
Delivery Date: | 15 |
Water Soluble Acid and Alkali Testing Instrument for Petroleum Products
Introduction:
GD-259 is designed and made as per the National Standard of People's Republic of China GB/T259 "Test Methods for Water Soluble Acid & Base of Petroleum Products". It is suitable to determine the water soluble acid & alkali in the liquid petroleum products, additives, lubricant greases, paraffin and waxy components as per GB/T259.
GD-259 extracts the water soluble acid & alkali in the sample using distilled water or ethanol solution. Then check color change of extracted solvent by methyl orange or phenolphthalein indicator respectively, or determine the pH value of extracted solvent using acidity meter to ascertain whether there exists the soluble acid and alkali in the extracted solvent.
Main technical features:
1. Simple design and meeting requirements of GB/T 259.
2. Heating power can be adjusted continuously. Easy to control.
3. Easy to operate. The result is accurate.
Main Technical Specifications:
1 | Power supply | AC 220 V±10%, 50 Hz |
2 | Heating power | 100 W~1000 W continuously adjustable |
3 | Acidity meter | Full scale: 0~14.00 pH |
Accuracy: ±0.01 pH | ||
4 | Separatory funnel | 250 ml |
5 | Graduated flask | 100 ml, 50 ml |
6 | Test tube | Φ18 mm×100 mm |
7 | Conical flask | 100 ml |
8 | Ambient temperature | -10 ℃~+35 ℃ |
9 | Relative humidity | ≤85% |
10 | Maximum power consumption | 1200 W |