Introduction

    GDGS different frequency automatic dielectric loss tester is a high-precision instrument, which can be applied to automatically measure electrical loss tangent ( ) and capacitance of all kinds of high voltage electrical equipments' on the spot of power plants, substations etc.

 

II.Technical Index


1.Condition of Use-5°C~40°C, RH<80%
2.Power SourceAC 110V±10%  Without Frequency Limit
3.High Voltage Output


Range1KV~10KV, every 1KV

Precision5%

Maximum Current140mA

Capacity1000VA
4.Self-Excited Power


Voltage RangeAC  0V~30V

Frequency45Hz /55Hz

Maximum Current15A
5.Definitiontgδ: 0.001%  Cx: 0.1PF
6.Precision:



7.Measuring Rangetgδ Without Limit




10kV
1kV

CVT Test
8.Size470mm(L) ×320mm(W)×390mm(H)

Weight30kg