Winner2000ZDE laser particle size eanalyzer adopt MIE scattering principle, which is high efficient-cost model with full automatic operation, adopts comprehensive scattering light detection system with highly sensitive ring photoelectric detector improving the test accuracy greatly.
Main Specifications:
Model Name | Winner2000ZDE | Winner2000ZD(3 Gears) |
Standard | ISO13320-1:1999, GB/T19077.1-2008, Q/JWN001-2009 |
Measuring Range | 0.1-300μm | 0.1-40μm /0.6-120μm/1-300μm |
Channels Number | 39 | 32×3 |
Accuracy | <1% (National Standard Sample D50 ) |
Repeatability | <1% (National Standard Sample D50 ) |
Light source | High performance He-Ne Laser (λ= 632.8nm, P>2MW) |
Dispersion Method | Ultrasonic | Frequency:40KHz Power:35W, Time: ≥1S |
Stir | Revolutions Speed: 0-300RPM (Adjustable) |
Circulate | Rated Flow:8L/min Rated Power:10W |
Sample Pool | Volume:350mL |
Micro- Sample Pool | Volume: 10mL ( Optional) |
Operation Mode | Full automatic and manual control, freely choose |
Optical Calibration System | Full automatic |
Test Speed | <2mins for each time |
Volume | L2200px×W975px×H1150px |
Net Weight | 41Kg |
Warranty | 2 years |
Main Features:
1,Advanced design of light path:
A patented technique of Fourier transform of converging light released the scattered light at large-scattering-angles from the restriction of the aperture of the Fourier lens. The focal length is reduced to enhance the resolution of the instrument.
2,Built-in dispersing units:
We carefully aligned the stirring set-up, the ultrasonic dispersing unit and the sample circulation pipes, and fixed them inside the instrument. Such a built-in design effectively prevents the inhomogeneous dispersion and sedimentation of big particles, which can be observed in the designs that these dispersing units are separated from the instruments, where the sample circulation pipes are therefore too long, The sample will be sufficiently dispersed.
3,Unconstrained fitting techniques:
The particle analysis software uses a unique unconstrained data fitting technique that we developed to obtain data of unknown size distribution, this is particularly important for researchers.
4,Micro sample chamber (optional):
The capacity of the sample chamber is as small as only 10ml. This helps with measuring expensive/precious samples, or samples difficult to be dispersed within medium.
5,Modern measurement control:
Users can perform all measurement procedures by simply operating on the PC and have ideal results in a very short time.
6,User-friendly Operation:
manual mode and the automatic mode, freely choose, to measure according to the sample features. In some conditions (e.g. the sample have unknown features or there are special requirements for the measurements), users can make a test measurement in the manual mode first, and after having an idea of the sample features and the measurement conditions, measure the samples in the automatic mode.
7,Fully automatic light path alignment:
A precise four phase hybrid stepping motor automatically aligns the optical path and can adjust it at any moment. This releases users from manual adjusting the optical path and improved accuracy and stability of the measurement results.
8,Quick measurements:
set " automatic" mode, all operation procedures are performed automatically,automatic water supply, automatic ultrasonic sample, stirring, circulation,background testing, sample testing,analysis , draining and cleaning, which significantly reduces the time for measurements, the full process only take 2 minutes.
9,Data analysis:
Errors in the data are rejected and the measurement results are automatically processed. Manual data processing is not necessary and the output is more standard.