user login

Drop ball impact test machine

2000.0~20000.0 USD
Min. Order: 1 Set/Sets
Trade Term: EXW
Payment Terms: T/T
Supply Ability: in stock
Place of Origin: Shandong

Company Profile

Location: Jinan, Shandong, China (Mainland)
Business Type: Manufacturer
Main Products: COF Tester, Oxygen Gas Transmisstion Rate Tester, Leak Tester, Gas Permeability Tester, Tensile Testing Machine

Product Detail

Means of Transport: Air
Usage: Impact Testing Machine
Power: Electronic
Brand Name: Labthink
Test method: Free Faling Dart Or Falling ball
Standards: ASTM D1709
Export Area: whole world
Production Capacity: in stock
Packing: veneer case
Delivery Date: 7 days
Show

Product Description

PRODUCT DESCRIPTION

FDI-01 Drop Ball Impact Tester/Testing Machine is applicable in the impact result and energy measurement of the falling dart from a certain height against plastic films and sheets with a thickness less than 1mm, which would result in 50% tested specimen failure.

 

CHARACTERISTICS

Micro-computer calculation, LCD display and Manu interface;

Two test methods integration: A & B.

Built-in observation light;

Pneumatic specimen clamp and release;

Falling dart electromagnetic suspension and automatic release;

Two start modes: Manual and foot-pedal;

Automatic testing status judgment;

Multi-units result display and printing;

Mini-printer;

RS232 interface;

 

PRINCIPLE

At the beginning of the test, choose test method, and estimate an initial mass and Δm. Start test. If the first specimen fails, decrease the mass of the falling dart by decreasing weight of Δm. If the first specimen is not a failure, add the mass of the falling dart by increasing weight of Δm. Test as the same. In brief, increase or decrease the Δm according to whether the former specimen is a failure or not. After 20 specimens, calculate the total number of failure specimens N. If N equals to 10, test over. If N is less than 10, add specimen and continue to test until N is equal to 10. If N is more than 10, add specimen and continue the test until the number of non-failure specimen is 10. Then the tester calculates the test results automatically.

according to special formula.

 

TECHNICAL INDEXES

Test Method

A, B (Optional)

Test Range

method A: 50~2000g

method B: 300~2000g

Accuracy

0.1g (0.1J)

Specimen Clamping

pneumatic

Gas Source Pressure

0.6MPa (Users provide gas source themselves)

Inlet Size

Φ8mm polyurethane pipe

Specimen Size

>150mm x 150mm

Power

AC 220V 50Hz

Net weight

70kg

 