Technical Date
L8P fetal/maternal monitor
Fetal Heart Rate:
Transducer: Multi-crystals, pulsed doppler , high sensitively
Strength: <5mW/cm2
Working frequency: 1.0MHz
Signal processing: special DSP system and modern recognization.
Measurement range: 50~210 bpm
Alarm Range:
High limit: 160,170,180,190 bpm
Low limit: 90,100,110,120 bpm
Maximum audio output power : 1.5 Watt
Toco:
Measurement range: 0~100 units
Maternal SpO2 measure:
Measure scope : 70%~99%.
Measure accuracy: ±3%
Pregnant HR Measure:
Measure scope: 30bpm~240bpm
Measure accuracy: ±2 bpm
NIBP measure:
a) NIBP-SYS (6.7~32.0Kpa)(50~240mmHg)
b) NIBP-DIA (3.4~26.6Kpa)(25~200mmHg)
c) NIBP-MEAN(2.0~24.0ka(15~180mmHg)
NIBP Accuracy: ±1.1kPa(±8mmHg)or ±5% of the results, take the bigger one.
Maximum Mean error : ±5mmHg
Maximum Standard deviation 8mmHg
Pulse Rate Accuracy: Maximum Mean error : ±2bpm
Measure accuracy: ±2 bpm or ±5% of the results, take the bigger one.
Measure mode: Manual
ECG
Heart Rate Measurement and Alarm Range: 15-300bpm
Accuracy: ±1bpm
Heart Rate Average: 8 beats
Lead Selection: II
Input: 5-lead ECG cable
Sweep Speed: 25 mm/s
Amplitude Selection: x1/4, x1/2, x1, x2, x4
Frequency Response:
Monitoring: 0.5 to 35Hz (+0.4dB, 3.0dB)
Defibrillation Protection: 5kV test
Recovery Time Following Defibrillation: Less than 5 seconds
Temperature:
Measurement and Alarm Limit: 0-50°C
Probe: Skin surface
Unit: Celsius/Fahrenheit
Accuracy: ±0.1°C(25°C ~45°C), ±0.2°C(0°C ~25°C or 45°C ~50°C)
Resolution: 0.1°C
Respiration
Measurement Mode: Thoracic Impedance
Respiration Rate: 0-150 bpm
Waveform Display Speed: 25mm/s
Display:
The LCD displays FHR, TOCO, FM , maternal parameters , time, date, volume and so on, It can store and play
back the data.
Power: a.c.100-240v, 50/60Hz or 16.8v Li-ION battery
Power Consumption: < 20W
Dimension: 350W×320D×85H(mm)
Net weight:3.5kg
Packing size:41.3×39-×24.5 (mm)
Gross Weight:9Kg
Environment
Working environment: Temperature: +5°C ~ +40°C
Humidity: < 85%
Atmospheric pressure: 86kPa ~ 106kPa
Transport and storage temperature: Temperature: -10°C ~ +55°C
Humidity: < 93%
Atmospheric pressure: 86kPa ~ 106kPa
Transducer Acoustic Output:
Under the requirements laid down in IEC 1157, 1992, the peak negative acoustic pressure
does not exceed 1Mpa. The output beam intensity does not exceed 20mW/cm2
and the spatial-peak temporal-average intensity does not exceed 100mW/cm2
GEL:
Viscous aqueous non-sensitizing , hypo-allergenic and non-irritating to skin. Indefinite shelf life,
Bacteriostat (not sterile).