user login

Three Zone Thermal Shock Test Chamber / High Low Temperature Testing

Place of Origin: Zhejiang

Company Profile

Location: Shenzhen, Guangdong, China (Mainland)
Business Type: Manufacturer
Main Products: Geophone Sensor, Geophone Strings, 3 Component Geophone

Product Detail

Model No.: HD-E703

Product Description

Three Zone Thermal Shock Test Chamber / High Low Temperature Testing

 

1. Description:

Three-Zone High-Low Temperature Thermal Shock Environmental Test Chamber occurs when a thermal gradient causes different parts of an object to expand by different amounts. This differential expansion can be understood in terms of stress or of strain, equivalently. At some point, this stress can exceed the strength of the material, causing a crack to form. If nothing stops this crack from propagating through the material, it will cause the object's structure to fail.

 

 

2. Specifications of Three-Zone High-Low Temperature Thermal Shock Test Chamber:

 

System

Two-zone test by means of damper switching

Three-zone chamber

Model

HD-E703-50A/HD-E703-50W

HD-E703-100A/HD-E703-100W

Customized

Performance

Test area

High temp. exposure range*1

+60~ to +200°C

Low temp. exposure range*1

-65to 0 °C

Temp. fluctuation *2

±1.8°C

Hot chamber

Pre-heat upper limit

+200°C

Temp. heat up time*3

Ambient temp. to +200°C within 30min

Cold chamber

Pre-cool lower limit

-65°C

Temp. pull down time*3

Ambient temp. to -65°C within 70min

Temp. recovery (2-zone)

Recovery conditions

Two-zone: High temp. exposure +125°C 30 min, Low temp. exposure -40°C 30 min; Specimen 6.5 kg (specimen basket 1.5kg) 

Temp. recovery time

Within 10 min.

Construction

External material

Cold-rolled rust-proofed steel plate

Test area material

SUS304 stainless steel

Door*4

Manually operated door with unlock button

Heater

Strip wire heater

Refrigeration unit

System*5

Mechanical cascade refrigeration system

Compressor<

Post Buying Request