Place of Origin: | Zhejiang |
---|
Location: | Shenzhen, Guangdong, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Digital Heating Thermostat, Fan Coil Thermostat, Heat Pump Thermostat |
Model No.: | T92+ UV-Vis Spectrophotometer |
---|
Visible UV-Vis Spectrophotometer with Thermostat battery photomultiplier tube
Spectral Bandwidth 0.1- 5.0nm Continuously Changeable, UV-Vis Spectrophotometer T92+
Applications:
Pharmaceutical production and quality control
Bio-tech & life science
Environmental monitoring & control
Food & Beverage production and quality control
Agricultural monitoring
Clinical & forensic analysis
Geology & metallurgy
Gas & oil
Specifications:
Optical system |
Double beam |
Scan speed |
Selectable |
Wavelength range |
190-900nm |
Wavelength accuracy |
+0.3nm |
Wavelength reproducibility |
≤0.1nm |
Spectral bandwidth |
Continuous slit 0.1- 5.0nm with 0.1nm interval |
Photometric mode |
Transmittance, Absorbance, Energy Concentration, all using UVWin software |
Photometric range |
-4.0-4.0Abs |
Photometric accuracy |
0.002A (0-0.5A), 0.004A (0.5-1A), 0.3%T (0-100%T) |
Photometric reproducibility |
0.001A(0-0.5A) |
Photometric noise |
0.0004A(500nm) 30min warm-up |
Baseline flatness |
0.001A(200-850nm) |
Baseline stability |
0.0008A/h (500nm, 0Abs), 2hr warm-up |
Stray light |
≤0.01%T(220nm NaI, 340nm NaNO2) |
Standard functionality |
No stand alone function |
Cell holder |
Fixed position sample and reference |
Detector |
Photo multiplier tube |
Light source |
Tungsten Halogen and Deuterium arc lamps |
Display |
No display |
Printer |
|