user login

Combined Flame / Graphite Furnace System Atomic Absorption Spectrophotometer PG990FG

Place of Origin: Zhejiang

Company Profile

Location: Shenzhen, Guangdong, China (Mainland)
Business Type: Manufacturer
Main Products: X-ray Diffractometer, Biology Lab Instruments, Pretreatment Device

Product Detail

Model No.: PG990FG Atomic Absorption Spectrometer

Product Description

Combined Flame and Graphite Furnace System PG990FG Atomic Absorption Spectrophotometer

 


Features & Functions

  • The flame and graphite furnace is integrated into one instrument.The change over from one technique to another is carried out by simple keystrokes within the software.
  • The instrument has a motorized 8 hollow cathode lamp turret which allows the automatic positioning and optimization of each hollow cathode lamp by the software.
  • The control of the gas flows for the fuel gas (C2H2) of the burner is also carried out directly from the software thus allowing optimization of the instrument for the analytical parameters for a selected analysis
  • Two methods of background correction are available. The first utilizes a Deuterium Arc lamp and the second is the proven method of Self Reversal.
  • High precision minimal optics ensures maximum light throughput to the computer controlled Czerny-Turner Monochromator.

Safety
 

  • Pressure monitoring for all gases 
  • Flame sensor
  • Drain trap level sensor
  • Gas leak detector 
  • Over pressurein premix
  • Safety cut off switch
  • Argon gas pressure sensor
  • Water flow sensor
  • Over temperature sensor
  • Broken graphite tube protection

Applications:

Agricultural
Clinical
Environmental
Food
Metal
Mining
Geological
Pertrochemical
Pharmaceutical

 

Specifications:

 

Optic System

Wavelength range

190-900nm

Monochromator 

Czerny-turner configuration

Bandwidth 

0.1, 0.2, 0.4, 1.0, 2.0, 5.0nm

Wavelength accuracy

+0.25nm

Wavelength repeatability

0.15nm

Baseline stability

0.005A/30min

Flame Analysis

Sensitivity (Cu)

0.03μg/ml/1% 

Burner heads

Titanium alloy burner

Nebulizer 

High efficiency glass

Atomization chamber

Corrosion-resistant material

Position adjustment

Automatic changeover of flame and furnace     Automatic setting of optimum height for flame burner

Safety 

Automatic ignition and of mixing air-acetylene gas with safety control

Graphite Furnace Analysis 

Character value(Cd)

0.5Pg

Temperature range

Ambient-2650

Heating 

Voltage feedback control when drying and ashing

Optical temperature control when atomizing

Heating program

Up to 9 steps with choices of ramp,temperature increase and full-power heating

Background correction

Deuterium lamp background correction

1.0Abs

Post Buying Request