user login

capacitive liquid pressure sensor

28.0 USD
Min. Order: 1 Piece/Pieces
Payment Terms: Paypal, L/C, D/P, D/A, T/T, WU, Money Gram
Supply Ability: 1000
Place of Origin: Shaanxi

Company Profile

Location: Xi'an, Shaanxi, China (Mainland)
Business Type: Manufacturer

Product Detail

Brand Name: Yunyi
Production Capacity: 1000
Packing: Standard
Delivery Date: within 7 days
Show

Product Description

Description :

Pressure or differential pressure will cause changing of capacitance value, analog signal convert to digital signal via AD and then being transfered to microprocessor, processed by microprocessor then convertted to 4~20mA analog electricity signal via DA. 

Features :

Super performance, use to measure pressure, differential pressure,liquid level, rate of flow.
Accuracy: ±0.075%~±0.1%F.S
Stability: ±0.25%
Standard 4~20mA,RS-485 communication interface
Support upgrade of basic scene control technique

Application
Compatible with 316LSS,noncorrosive liquids.   
Shipping and aviation industry.
Hydraulic pressure and pneumatic control system
Apply in Petroleum industry, chemical industry, electric power, metallurgy, hydrology etc process control.

Range

Range code

Range

Measuring range

Min.

Max.

4

0.374KPa

37.4KPa

-37.4~37.4KPa

5

1.868KPa

186.8KPa

-186.8~186.8KPa

6

6.9KPa

690KPa

-690~690KPa

7

20.68KPa

2068KPa

-2068~2068KPa

Technical Data

Output

Analog output: two –wire 4-20mADC

Digital output: two –wire 4-20mADC and superimposing variable digital signal accord with HART

Power supply

supply voltage 12.5-45VDC

Limit load

Max load resistor Rmax in loop decided by the supply voltage, see the below drawing. HART PROTOCOL require the load resistor not less than 250Ω,not more than 600Ω

Communication distance

Max. 1.5Km, multi-address communication can connect 15 smart transmitters Max.

Display

4 digital display,5 digit LCD display, pointer type.

Explosion-proof

ExiaIICT5;explosion suppression type

Zero offset

Can positive transfer or negative transfer, but can not overstep the utmost value

Temperature

ambient temperature -40-85℃(When choose LCD : -20-70℃)

Medium temperature

-40~104℃

Static pressure and single side overpressure

3.45KPa(absolute pressure)~31.2MPa(gauge pressure)

Relative humidity

5~98%RH

Run-up time

less than 5 sec.

Capability change

Less than 0.16cm3

Damp adjust

0.20~25 second

Malfunction alarming

If occur malfunction, automatism diagnose function will output 3.9 or 21.0mA

Configuration protection

To avoid the datas are modified random, can utilize software or protect switch lock the record data


Post Buying Request