Place of Origin: | Jiangsu |
---|
Location: | Shenzhen, Guangdong, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Medical Hospital Beds, Pediatric Hospital Bed, Hospital Bed Accessories |
Model No.: | AYR - HL25 |
---|
- 25 Degree Hospital Low Temperature Medicine And Viccne Storage Freezer
Quick Detail :
1 . Hospital refrigeator
2 . laboratory Freezr
3 . Ultra low temperature freezer
Specifications :
Product mode | Total useful volume | Temperature inside the box |
Exterior size (Width * Depth * Height) |
AYR - HL25 | 120L | - 10 - 25 | 685 × 530 × 950mm |
AYR - HL25 | 170L | - 10 - 25 | 950 × 530 × 950mm |
AYR - HL25 | 200L | - 10 - 25 | 1050 × 530 × 950mm |
AYR - HL25 | 250L | - 10 - 25 | 1150 × 630 × 950mm |
AYR - HL25 | 300L | - 10 - 25 | 1150 × 720 × 950mm |
AYR - HL25 | 460L | - 10 - 25 | 1750 × 700 × 930mm |
AYR - HL25 | 560L | - 10 - 25 | 1780 × 780 × 980mm |
Detail Description :
1 . Adopt the original binding Germany DANFOSS high-performance compressor |
2 . Adopt the original binding “Bi-Sonic” double fan strongly against the low temperature and moisture |
3 . Adopt the Germany “EBM” heat elimination fan |
4 . Adopt the Japanese Matsushita vaporizer (cooling pipeline inside the box) |
5 . Fluoride-free refrigeration system for environmental protection |
Optional Accessories :
The EPS emergency supply |
The DC24V-AC220V inverter, and the DC24V power supply can keep the refrigerator under normal operation. |
A24-hour follow –up reading and printing system(option ) |
Precise temperature control system
1 . The high - accuracy computer - controlled temperature system can maintain the temperature constancy in the box at 4 ± 1 ° C precisely . The temperature set - value can be adjusted at 0 ° C ~ 6 ° C , and the temperature display precision is 0 . 1 ° C , the resolution is 0 . 1 ° C and the adjusting increment is 0 . 1 ° C .
2 .The system has special high - resolution temperature calibrating function . That means the set - value figure can be displayed and calibrated at 0 ~ 14 ° C . The calibrating increment is 0 . 1 ° C .
3.The forced cold - air circulation system with intelligence - control fan can ensure the temperature uniformity inside the box .
4 . Keyboard lock , password protection function , and method against missing - adjustment of the operating parameter .
5 . Setting of the alarm temperature range .
6 . 7 system failure alarms ( high temperature alarm , low temperature alarm , sensors failure alarm , door - opening alarm , electricity failure alarm , low back - up battery alarm , low voltage alarm ) .