Laboratory Flammability Testing Equipment , IEC60695-11-5 Needle Flame Test Apparatus

Place of Origin: Zhejiang

Company Profile

Location: Shenzhen, Guangdong, China (Mainland)
Business Type: Manufacturer
Main Products: Flammability Testing Equipment, Cable Testing Equipment, Toys Testing Equipment

Product Detail

Model No.: NF-1

Product Description

Laboratory Testing Equipment IEC60695-11-5 Needle Flame Test Chamber

 

 

Product information:

The Needle Flame Tester consists of a 0.9mm needle-shaped burner that tilts with the angel of 45°and provide it with butane gas. The fire hazard of the equipment when there is flammule inside the equipment that caused due to the fault condition is assessed by whether the specimen can be ignited or the combustion time and length. This tester conforms to the standard requirement of IEC60695-11-5,IEC60335-1.

The Needle Flame Tester is suitable for the research, production and quality testing department of lighting equipment, low-voltage apparatus, household appliance, machine tool electric appliance, electrical machine, electric tool, electronic instrument, electrotechnical instrument, electrical connector, electrical and electronical equipment and other components. It is also suitable for insulating material, engineering plastics or other solid combustion material industry.

 

Model specification and technical parameters (Optional for touch screen NF-2):

 

Parameter/ model

NF-1

Working voltage

220V/50Hz

Timing device

9999X0.1S

Combustor

Inner boreφ0.5mm ± 0.1mm ,OD ≤φ0.9mm ,L ≥ 35mm,stainless steel

Combustion angle

tilt 45°( testing ) or vertical ( when adjust the height of the flame)

Height of the flame

12±1mm adjustable(with flame height gauge)

Gas category

 95% butagas ( can use the bottled lighter gas)

Range of temperature test

0~1000

Flame warming requirement

From 100°to 700°,time for standard values within 23.5S±1S

Temperature thermocouple

φ0.5 armour nickel-chromium/nickel aluminum wire K-Thermocouple

Work shop

0.5 cube,staunless steel case,black yogon background,(customizable0.75 cube,1cube)

Testing process

Test procedure automatic control,independent air draft

Volume size adjustment

Flowmeter adjustment

Conforms  to standard

GB4706.1-2005,GB5169.5-2008,IEC60695-11-5,IEC60335-1

 

Includes:
 A white Pine Test Board and five Sheets of Tissue Paper
 A flame height gauge
 Consumables:
 White Pine Test Board and Tissue Paper
 Thermocouple Type K and Copper Block