Min. Order: | 1 Piece/Pieces |
---|---|
Payment Terms: | T/T, WU |
Supply Ability: | 1000 pcs/month |
Place of Origin: | Jiangsu |
Location: | Suzhou, Jiangsu, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Hot Runner Temperature Controller, Temperature Humidity Controller, Temperature Humidity Transmitter, Temperature Sensor |
Model No.: | TKSD |
---|---|
Means of Transport: | Ocean, Air |
Brand Name: | Tinko |
Power supply: | DC or AC 24V±10% |
Measurement range: | H: 0%RH~100% RH T: 0℃~50℃ |
Accuracy: | H: ±3%RH(5%RH~95%RH,25℃) T: ±0.5℃(0℃~50℃) |
Working temperature: | -10℃~60℃ |
Long-term stability: | H:<1%RH/year T:<0.1℃/year |
Response time: | <15s(1m/s wind speed) |
Output signals: | DC.4mA~20mA |
Load capability: | ≤500Ω |
Production Capacity: | 1000 pcs/month |
Packing: | Standard Carton Packaging |
Delivery Date: | Within 7 days after receive payment |
Introduction:
TKSD digital Temperature Humidity sensor which Adopt integrated design is widely used for measureing in HVAC.It has import search unit,intelligent core structure,and more output signals selectable. the sensor with dependable performance, long perfomance life, fast Response time.and ,it is used with temperature &humidity controllers to monitor and control the temperature and humidity.
Application area:
Widely used for communication room,office room,workshop,storehouse,hospitals,,monument room,museum,refrigeratory,HVAC and so on.also used on the machine of humidification or dehumidification.
Features:
1)Response time:<10s
2)Sensor and probe:0 ~50 ℃, 0 to 85% RH
3)Working temperature:-10~60℃
4)Output signal:4-wire 4~20mADC /0~10V/RS485 (optional)
5)Load Capacity:Current:<500ω voltage:="">3KΩ
6)Cable:PG7 cable gland
7)Installation Method:Duct mounting
8)Shell Size: housing:ABS100×68×50mm;probe+sleeve 12×190mm
9)Accuracy: Grade A.±0.3℃(0~50℃), ±2%RH(5%RH~95%RH,25℃);
Grade B.±0.5℃(0~50℃), ±3%RH(5%RH~95%RH,25℃)
10)Long-term stability: Humidity<1% RH / year
Temperature<0.1 ℃ / year
11) Range:T:0~50℃or -40~120℃,H:0%~100%RH
Selection Table:
TKSD | Duct mounting with flange | ||||||
-2 | Temperature & Humidity Transmitter | ||||||
-3 | Temperature Transmitter | ||||||
8 | 4-wire 4~20mA output | ||||||
9 | 4-wire 0~10V output | ||||||
N | No output | ||||||
C1 | RS485 output | ||||||
N | No RS485 output | ||||||
L | With display | ||||||
N | No display | ||||||
W | wide range(-40~120℃) | ||||||
N | 0~50℃ | ||||||
A | High accuracy |