user login

Tinko multichannel temperature scanner

Min. Order: 1 Piece/Pieces
Payment Terms: T/T, WU
Supply Ability: 500 pcs/month
Place of Origin: Jiangsu

Company Profile

Location: Suzhou, Jiangsu, China (Mainland)
Business Type: Manufacturer
Main Products: Hot Runner Temperature Controller, Temperature Humidity Controller, Temperature Humidity Transmitter, Temperature Sensor

Product Detail

Model No.: CTM-1X
Means of Transport: Ocean, Air
Brand Name: Tinko
Power supply: AC85~265V (DC24V optional)
Measurement accuracy: ±0.5% F.S.
Communication: RS485, Modbus RTU (optional)
Micro-printer drive: TTL (optional)
Working circumstance: -10~50℃, 20~80%RH
Panel size: 160*80mm
Mounting panel cutout size: 150*76mm
Production Capacity: 500 pcs/month
Packing: Standard Carton Packaging
Delivery Date: Within 7 days after receive payment
Show

Product Description

Application:


The scanner can be used to monitor signals changed slowly or response speed undemanding occasions, such as temperature and humidity of telecommunication room, office, workshop, warehouse, hospital, library, etc.


Features:

1. Input signal type of each channel can be selected (RTD, thermocouple, current and voltage are not universal)

2. Channel parameters setting can be unified or separate

3. 32 alarm points with indicator, each channel can be set one or more alarm point

4. 2 alarm relay outputs with indicator, they can be shared by all alarm points or used by one

5. It can monitor multichannel one by one or one selected channel

6. Number of active channels and switching time can be set

7. 2 transmitting outputs can be set minimum or maximum or average value of all channels, or value of selected channel

8. Measurement value can be corrected

9. Real time clock, it can drive micro-printer to print each channel value, maximum & minimum & average value of all channels and present time

 

Specification:

1. Input signal (1~16 channels): RTD: Pt100 (-199.9~599.9℃),

                                 Cu50 (-49.9~149.9℃)

                                 Thermocouple: K (-50~1300℃), E (-50~1000℃),

                                 J (-50~1000℃), N (-50~1300℃), T (-50~400℃),

                                 B (-50~1800℃) , S (-50~1600℃), R (-50~1700℃)

                                 DC current: 4~20 mA, 0~20mA (-1999~9999 free )

                                 DC voltage: 1~5V, 0~10V (-1999~9999 free)

2. Measurement accuracy: ±0.5% F.S.

3. Display unit:Measurement value: 4 digital 1.0"LED

                Channel number: 4 digital 0.36"LED

4. Alarm:  2 alarm relay NO contacts (AC250V, 3A )

           8 alarm modes, selected by each channel

           32 alarm points, free combination with channels & alarm relays

5. Transmitting output: DC.4~20mA, DC.0~20mA

  (5 transmitting modes selectable) (optional)

6. Communication: RS485, Modbus RTU (optional)

7. Micro-printer drive: TTL (optional)

8. Power supply: AC85~265V (DC24V optional)

9. Working circumstance: -10~50℃, 20~80%RH

10. Panel size: 160mm*80mm

11. Mounting panel cutout size: 150mm*76mm



Post Buying Request