Payment Terms: | L/C, D/P, D/A, T/T |
---|---|
Place of Origin: | Chongqing |
Location: | Chongqing, China (Mainland) |
---|---|
Business Type: | Manufacturer, Trading Company |
Means of Transport: | Ocean, Air, Land |
---|---|
Usage: | Others |
Power: | Electronic |
Brand Name: | gold |
Packing: | wooden case |
GD-261-1 Semi-auto PMCC Flash Point Tester
The close cup flash point tester is designed and made as per the National Standard GB/T261 "Test Methods for Flash Point of Petroleum Products (Closed Cup Methods)".
The close flash point tester adopts microprocessor techniques and equipped with advanced software, it can control temperature, stir sample, ignite, detect, and print out test data automatically. It has a six digits display.
Main parameters
(I) Operation condition
1. Ambient temperature: +10℃~+30℃;
2. Relative humidity: < 80%;
3. Power supply: AC 220V±10%, 50±1Hz;
(II) Temperature measurement
1. Measurement range: ambient temperature~250℃;
2. Accuracy: 0.2%;
3. Resolution: 0.1 ℃;
4. Repeatability:
(1) Flash Point is lower than 104 ℃: error is 2 ℃;
(2) Flash Point is higher than 104 ℃: error is 4 ℃;
(III) Basic parameters
1. Interval for flame application
(1) Flash point is lower than 104 ℃: 1 ℃;
(2) Flash point is higher than 104 ℃: 2 ℃;
2. Temperature rising rate: as per GB/T261-83;
3. Stirring speed: 105 RPM
4. Ignition mode: by gas flame (4 mm);
5. Total power consumption: not more than 300 W;
6. Dimension:
(1) Main case: 360×110×300 (mm);
(2) Measurement case: 320×260×200 (mm);
7. Net weight: 10 kg;