Min. Order: | 1 Set/Sets |
---|---|
Trade Term: | FOB,CFR,CIF,DAP,EXW |
Payment Terms: | L/C, T/T, WU |
Supply Ability: | 100pcs/month |
Place of Origin: | Chongqing |
Location: | Chongqing, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Oil Purifier, Oil Testing Equipment, Electrical Testing Set |
Model No.: | GD-265D-1 |
---|---|
Means of Transport: | Air |
Usage: | Auto Testing Machine |
Power: | Electronic |
Brand Name: | Gold |
Testee: | Liquid petroleum products |
Display: | Digital Display |
Bath: | Double Bath |
Test Sample: | 4 Samples one time |
Temperature controlling accuracy: | ±0.01℃ |
Timing device: | 0.1s -999.9 s |
Resolution: | 0.1s |
Warranty: | 1 Year |
Production Capacity: | 100pcs/month |
Packing: | Wooden Case |
Delivery Date: | 7~10 days |
Summary
This instrument is designed and made as per the industry standard of People's Republic of China SY/T5651 Technical Condition of Petroleum Products Kinematic Viscosity Tester, GB/T 265 Petroleum products-Determination of kinematic viscosity and calculation of dynamic viscosity and ASTM D445 Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids. It is suitable to determine kinematic viscosity of liquid petroleum products (Newtonian liquids) by measuring the time for a volume of liquid to flow under gravity through a calibrated glass capillary viscometer at a constant temperature.
I. Main technical features
1. The instrument is a kind of specially-made testing machine which the temperature control accuracy can reach ±0.01℃ and having digital display.
2. It adopts hard glass bath and heat preservation cover(double shell).Good heat prevention and easy to observe the sample.
3. It adopts desktop and all-in-one machine design,convenient to use.
4. It adopts electric stirrer to ensure the uniform of bath temperature.
II. Main technical specifications
1. Power supply: AC 220 ± 10 %, 50 Hz.
2. Heating power:1000W(auxiliary heating), 600W(temperature control heating)
3. Power for stirrer: 6W, rotating speed:1200RPM
4. Temperature range: room temperature~100 ℃.
5. Temperature control accuracy: ±0.01℃.
6. Timer: 0.0s~999.9s
7. Constant temperature bath: 20L, double shall structure.
8. Ambient temperature: -10℃~+35℃
9. Relative humidity: <85%.
10.Maximum power consumption: 1800W.
11.Capillary viscometer tubes(Pinkevitch viscometer): 7 pieces in total. The inner diameters for each: 0.6, 0.8,1.0, 1.2, 1.5, 2.0 and 2.5mm
12. Dimension: 530mm×400mm×670mm
Net weight: about 42Kg
Note: This instrument is equipped with 7 pieces of Pinkevitch viscometers.We can customize instrument equipping with Ubbelohde viscometers according to customer's requirements.