10.4 Screen Auto Hematology Analyzer (3 Diff 22 Parameters)

3200.0 USD
Min. Order: 1 Piece/Pieces
Trade Term: FOB
Payment Terms: Paypal, L/C, T/T, WU, Money Gram
Supply Ability: 150 per Month
Place of Origin: Henan

Company Profile

Location: Zhengzhou, Henan, China (Mainland)
Business Type: Manufacturer, Trading Company, Distributor/Wholesaler

Product Detail

Model No.: HP-HEMA6100
Means of Transport: Ocean, Air, Land
Flow cytometry: XP operation system
Closed Puncture sampling: avoid cross-infection
ISO,CE: 3-diff
22parameters: auto hematology analyzers
Production Capacity: 150 per Month
Packing: standard export package
Delivery Date: about 10 days
Show

Product Description

HP-HEMA6100  10.4 Screen Auto Hematology Analyzer (3 Diff, 22 Parameters)

The name of the machine is HP-HEMA6100 Automatic Hematology Analyzer,which mainly used to detect the parameters of the series of red blood cells and hemoglobin, platelet and leukocyte in the blood,It is a vitro diagnostic equipment which is applied to quantitative analysis of blood cells of clinical and laboratory.

The machine is mainly used for the quantitative measurement of the blood in the 19 parameters and 3 histograms:

Haematology Analyzer

Fully automated to increase productivity

3-part differential cell counter with 22 parameter and laser light source

Direct tube sampling from original collection vials for operator safety

Bar code capability to reduce manual input

Proven technology to provide accurate results

More than 60 samples per hour

Minimum maintenance required

 

General Information:                                                                           

Analyzer Packing Dimension(mm):530*466*595mm,  NW:20kg, GW: 24kg

Free Reagents packing : 360*300*280mm, NW:7KG, GW: 8KG  

Model

Screen size (inch )

Puncture

Storage

Graphical Touch Screen

Power

HP-HEMA6100

10.4

None

20K

NO

50Hz±3Hz

 

Specifications

Test Items

WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram

Background Count

WBC: ≤0.3×109/L   RBC: ≤0.03×1012/L   HGB: ≤1g/L   PLT: ≤10×109/L

Linearity

Test Item

Test Range

Allow Difference of Linearity

WBC

0∼6.0×109/L

±0.3×109/L

6.0∼99.9×109/L

±5%

RBC

0∼0.99×1012/L

±0.05×1012/ L

1.0∼9.99×1012/L

±5%

HGB

0∼99 g/L

±2g/L

100∼300 g/L

±2%

PLT

0∼99×109/L

±10×109/L

100∼999×109/L

±10%

Repeatability

Test Item

CV(%)

Teat Range

WBC

≤2.0

7.0~15.0×109/L

RBC

≤0.5

3.5~6.0×1012/L

HGB

≤1.5

110~180g/L

PLT

≤4.0

150~500×109/L

MCV

≤0.5

80~110fL

Sample Volume

Twig blood: 20μl

Dilution Ratio

WBC/HGB: 1:500   RBC/PLT: 1:50000

Interfaces

RS232 joint, USB, PS2, Parallel port

Test Speed

60 samples/hours

Continuous Work Time

>8h

Note:More information for our products and service,please turn to our official website or contact us directly.

 

 

 

 