Min. Order: | 1 Piece/Pieces |
---|---|
Trade Term: | FOB |
Payment Terms: | Paypal, L/C, T/T, WU, Money Gram |
Supply Ability: | 150 per Month |
Place of Origin: | Henan |
Location: | Zhengzhou, Henan, China (Mainland) |
---|---|
Business Type: | Manufacturer, Trading Company, Distributor/Wholesaler |
Model No.: | HP-HEMA6100 |
---|---|
Means of Transport: | Ocean, Air, Land |
Flow cytometry: | XP operation system |
Closed Puncture sampling: | avoid cross-infection |
ISO,CE: | 3-diff |
22parameters: | auto hematology analyzers |
Production Capacity: | 150 per Month |
Packing: | standard export package |
Delivery Date: | about 10 days |
HP-HEMA6100 10.4 Screen Auto Hematology Analyzer (3 Diff, 22 Parameters)
The name of the machine is HP-HEMA6100 Automatic Hematology Analyzer,which mainly used to detect the parameters of the series of red blood cells and hemoglobin, platelet and leukocyte in the blood,It is a vitro diagnostic equipment which is applied to quantitative analysis of blood cells of clinical and laboratory.
The machine is mainly used for the quantitative measurement of the blood in the 19 parameters and 3 histograms:
Haematology Analyzer |
Fully automated to increase productivity |
3-part differential cell counter with 22 parameter and laser light source |
Direct tube sampling from original collection vials for operator safety |
Bar code capability to reduce manual input |
Proven technology to provide accurate results |
More than 60 samples per hour |
Minimum maintenance required |
General Information:
Analyzer Packing Dimension(mm):530*466*595mm, NW:20kg, GW: 24kg
Free Reagents packing : 360*300*280mm, NW:7KG, GW: 8KG
Model | Screen size (inch ) | Puncture | Storage | Graphical Touch Screen | Power |
HP-HEMA6100 | 10.4 | None | 20K | NO | 50Hz±3Hz |
Specifications | |||
Test Items | WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC Histogram, RBC Histogram, PLT Histogram | ||
Background Count | WBC: ≤0.3×109/L RBC: ≤0.03×1012/L HGB: ≤1g/L PLT: ≤10×109/L | ||
Linearity | Test Item | Test Range | Allow Difference of Linearity |
WBC | 0∼6.0×109/L | ±0.3×109/L | |
6.0∼99.9×109/L | ±5% | ||
RBC | 0∼0.99×1012/L | ±0.05×1012/ L | |
1.0∼9.99×1012/L | ±5% | ||
HGB | 0∼99 g/L | ±2g/L | |
100∼300 g/L | ±2% | ||
PLT | 0∼99×109/L | ±10×109/L | |
100∼999×109/L | ±10% | ||
Repeatability | Test Item | CV(%) | Teat Range |
WBC | ≤2.0 | 7.0~15.0×109/L | |
RBC | ≤0.5 | 3.5~6.0×1012/L | |
HGB | ≤1.5 | 110~180g/L | |
PLT | ≤4.0 | 150~500×109/L | |
MCV | ≤0.5 | 80~110fL | |
Sample Volume | Twig blood: 20μl | ||
Dilution Ratio | WBC/HGB: 1:500 RBC/PLT: 1:50000 | ||
Interfaces | RS232 joint, USB, PS2, Parallel port | ||
Test Speed | 60 samples/hours | ||
Continuous Work Time | >8h |
Note:More information for our products and service,please turn to our official website or contact us directly.