portable operating light
1. Product's mainly structure and performance characteristics
This product is applicable to ENT, urology, and gynecologic surgery lighting or auxiliary lighting, and it is the best lighting equipment. With high brightness, portability and other characteristics, it is more suitable for out-patient operating room of the emergency operations.
With the mobile design, Φ160mm large reflective masks, its lamp height and angles can be adjusted to meet the needs of any angle; its illumination can be achieved to 35000LUX. So it is an ideal lighting equipment for out-patient surgery.
2. Scope, Specifications Model
LED200 is used for the medical unit for surgical lighting
3. Technical parameters
3.1 Environment of the use
a)Scope of the environment temperature:10~50°C;
b)Scope of Relative Humidity:40~60%;
c)Scope of atmospheric pressure:860~1060hPa;
d)Power:~220V,50Hz.
3.2 Parameters table
1 Illumination 35000Lux
2 Color temp 4000±500K
3 Diameter of spot Φ50mm
4 Illumination depth ≥350mm
5 bulbs Cold source LED bulb
6 *Rechargeable battery support time 3 hours (*Optional configuration)
7 Max Height 2.00 meters
4. Maintenance and Repair
Check shadowless lamp connecting parts of the fastener of the state every two months
Check shadowless lamp circuit is good or not every two months
Check the flexibility of each activity of shadowless lamp parts every two months
Check the reliability of light bulbs and light bulb seat every two months
5. Storage and transport
Shadowless lamp should be stored in no more than 80% relative humidity, temperature 0 °C ~ ± 50 °C, and non-corrosive gases and well-ventilated room.
Shadowless lamp is fragile, vulnerable to sharp or hard objects, scratches or damage, please use the original packaging during transport, and transport operators by experienced transportation agents.
6. The time of safe use
This product is safe for 10 years (from the production date)