user login

Automatic oil Tester Equipment /Closed Cup Flash Point Apparatus/Automatic Pensky-Martens closed cup flash point

Place of Origin: Chongqing

Company Profile

Location: Chongqing, China (Mainland)
Business Type: Manufacturer
Main Products: Oil Tester , Electric Tester, Oil Purifier

Product Detail

Model No.: GD-261D

Product Description

Detail Informtion

 

GD-261D  Closed Cup Flash Point Apparatus

 

Purpose

 

The  Closed Cup Flash Point Apparatus is designed and made as per the National Standard of People's Republic of China GB/T261 "Test Methods for Flash Point of Petroleum Products (Closed Cup Methods)". It is suitable to determine closed cup flash point of petroleum products, which is described as the lowest temperature at which the mixture of a petroleum product sample vapor and air contacts flame and makes a flash when the sample is heated in a closed cup under the stipulated condition.

 

 

 Main Technical Specification and Parameters

 

Power supplyAC(220±10%)V, 50Hz)
Temperature measurementMeasurement range: room temperature~300°C
Repeatability: ≤2°C
Reproducibility: ≤4°C
Resolution: 0.1 °C
Accuracy: 0.5%
Basic parametersTemperature rising speed: as per the GB/T261
Ignition mode: by an electron igniter; gas flame for flame application
Ambient temperature10~40°C
Relative humidity ≤8°C
Total power consumptionnot more than 300 W.
Overall dimension520mm×360mm×310mm

 

Main technical characteristics

 

1. It adopts colored LCD, man-machine conversation interface, and blank buttons. It has prompts menu and guide to preset temperature, sample serial number, barometric pressure, test data and other parameters.

 

2. It can simulate and show temperature-time curve. It has prompts for wrong operation and prompts to modify test data, test time, and other parameters.

 

3. It can save test data of 100 groups. It is equipped with RS-323, and 485 computer ports, so it can save more test data when it is connected to a computer. Then you can also modify some parameters.

 

4. It can calibrate the effect of barometric pressure to the test and calculate the modified value.

 

5. Differential detection. It can modify system deviation automatically.

 

6. It can open lid, apply flame, make detection, and print test data automatically. The test arm can be raised up and lowered down automatically.

 

7. It uses an electron igniter for flame ignition. It has forcible wind cooling function.


Post Buying Request