user login

Medical ventilator, transport ventilator, emergency ventilator

1.0~2.0 USD
Min. Order: 1 Piece/Pieces
Trade Term: FOB
Payment Terms: T/T, WU
Supply Ability: 100000
Place of Origin: Jiangsu

Company Profile

Location: Taizhou, Jiangsu, China (Mainland)
Business Type: Manufacturer
Main Products: Medical Ventilator, Portable Ventilator, Transport Ventilator, Emergency Ventilator, Medical Device

Product Detail

Model No.: JIXI-H-100A
Means of Transport: Ocean, Air, Land
Production Capacity: 100000
Packing: carton box
Delivery Date: 7 days after payment
Show

Product Description

Use for first aid and ICU,it can be use for pediatric and adult patients
Physical Specifications
Dimensions
Width:    50cm             Depth:    50cm          
Height:   108cm          Weight    25Kg
Packing Dimensions
Width:    58cm             Depth:    55cm          
Height:   122cm          Weight    45Kg
Environmental Specifications
Operating Requirements
Temperature:            5 to 40 ℃;
Relative Humidity:        less than 80%;
The atmosphere pressure: 70 to 106 kPa ;
Storage/Shipping Requirements
Temperature:            -20 70℃
Relative Humidity:        less than 93%
The atmosphere pressure: 70 to 106 kPa
Electrical Specifications
Input Voltage Requirements
AC Voltage:        220V+/-10%
Frequency:     50HZ+/-2%
Power:             170 VA;
Fuse Specification Information
Quantity:         2
Current Rating:   0.8 A
Voltage Rating:   250 Volts
Type:            5*20 mm RT
Output Voltage Specification Information
AC Voltage:        220V+/-10%,
Frequency:      50HZ+/-2%
Power:             130VA;
Pneumatic Specifications
Oxygen Inlet
Pressurized Oxygen: 0.45 MPa +/-10% medical oxygen.
IEC Classification
Ordinary equipment, not protected against the ingress of liquids
Not protected/Not suitable for use in the presence of flammable
Anesthetic gases.
ClassⅠ, Type B
   Performance Specification And Tolerances
Ventilator
Parameters
Ranges
Remark
Controls                                                      
Power
ON/OFF
Modes
     IPPV (CMV)     IPPV+SIGH    SIPPV (A/C)  SIMV   MANUAL BREATH
Tidal Volume
0~1800ml
Breath Rate
5~60 bpm
PEEP
0~2KPa
%O2
45~90%
Sigh
Inspiratory Time
0.5~3sec
Exp. Hold Time
0~1sec
Trigger Level
-1~0kpa
Over Pressure Relief
≤6kpa
Alarm Set
Monitors                                                       
Tidal Volume
0~2560ml
MIN. Volume
0~99L
Total Breath Rate
0~99 bpm
%O2
0~99%
Main Circuit Pressure
-2~10KPa
Performance Specification And Tolerances
Ventilator
Parameters
Ranges
Tolerances
Alarms and Indicators                                            
High Pressure Alarm Limit
2~6kpa
flash, red indicator
+/-0.1kpa
Low Pressure Alarm Limit
0. 2~2kpa
Red indicator
+/-0.1kpa
High Minute Volume Alarm Limit
10~30L
flash, Red indicator
+/-5%
Low Minute Volume Alarm Limit
1~10L
flash, Red indicator
+/-5%
High Rate Alarm Limit
30~59bpm
flash, Red indicator
+/-2.5%
Apnea interval
15, 30, 45, 60second flash, Red indicator
+/-1second
Alarm Silence
60 sec. max
Power
Red indicator
Insp. phase
Green indicator
Exp. phase
Green indicator
IPPV Mode
Green indicator
IPPV+SIGH
Green indicator
SIPPV
Green indicator
SIMV
Green indicator
MANUAL
Green indicator
What Is Provided
Parts
Number
Specification
Ventilator
1
Mobile
1
Humidifier
1
Breath Circuit
1
Manipulators
1
Oxygen Pressure Reducer
1
Mask
1
Hand Book
1
CD For Operation
1
Post Buying Request