user login

Class N STE-12L-A Series A Steam Sterilizer Thermal Vacuuming Autoclave Class N

Min. Order: 2 Piece/Pieces
Trade Term: FOB
Supply Ability: 300 units a month
Place of Origin: Zhejiang

Company Profile

Location: Jinhua, Zhejiang, China (Mainland)
Business Type: Manufacturer
Main Products: Steam Sterilizer, Intra-oral Camera, Water Distiller, Ultrasonic Cleaner, Dental Consumables

Product Detail

Model No.: Class N STE-12L-A
Means of Transport: Ocean, Land
Brand Name: Tinget
Voltage(V/Hz): 110~240V/50-60Hz
Power(W): 1600 W
Chamber Size(mm): Φ200mm*360mm
Packing Size(mm): 690*560*490
Net Weight(kg): 38kg
Gross Weight(kg): 43kg
Color: White body and blue door handle
Certificates: CE,CMDCAS,ETL,ISO9001,ISO1348
Material: Stainless steel and plastic
Display: Digital Display
Production Capacity: 300 units a month
Packing: CTNS or wooden box
Delivery Date: 7 days after payment delivery
Show

Product Description

  It has a screen that allows you to view the pressure, temperature and the status of the autoclave during the sterilization cycle.

Its appearance is a little different from Class N STE-8L-A from the same series of series A steam sterilizer.

It has a clear and nice appearance which the body is pure white with a sky blue door handle,it is made of stainless steel and plastic which are of high quality.
Class N STE-12L-A Series A Steam White and Blue Sterilizer
Following is the graph of working cycle with Class N system:
Class N STE-12L-A Series A Steam White and Blue Sterilizer

Application Area:
Dental Clinics, Human Hospitals, Laboratory, Veterinary, Beauty Salon, Tattoo.


Features:

.Class N autoclave, without vacuum pump.

.LCD display.

.A built-in independent steam generator, making the sterilization fast and completely.

.Most of the main parts for TINGET autoclaves are imported from global famous company.

.Sterilization temperature: 121℃/1.1bar, 134℃/2.1bar.

.Open-type water storage tank which is very easy for filling and cleaning.

.Automatic water filling and draining.


Safety:

.Door protection locking system , prevents the cycle from starting unless the door is properly locked, and also prevents the door from opening if there is pressure in the chamber.

.Pressure safety valves prevent over pressurization in chamber and generator.

.A safety fuses prevent over heating in the chamber and generator.

.Able to detect and identify the exact cause of any problem by warning and showing a specific error code.

.Water level switch in the water tank, controlling the max and min water level.


Documentation:
.Have printer port, able to be connected TINGET printer to print the sterilization records for each cycle.

Program:7 programs

ProgramTemperature(℃)Press(Par)Sterilization Time(min)Total Time(min)
Solid1342.1418-30
Solid1211.12030-45
Wrapped1342.1630-40
Wrapped1211.12035-50
Cotton1342.11045-65
Cotton1211.13050-75
Plastic1050.23050-65













































Post Buying Request