Min. Order: | 1 Piece/Pieces |
---|---|
Trade Term: | EXW |
Payment Terms: | T/T, WU |
Place of Origin: | Guangdong |
Location: | Hong Kong |
---|---|
Business Type: | Manufacturer, Distributor/Wholesaler |
Model No.: | Model BD-6000 |
---|---|
Means of Transport: | Ocean, Air |
Overall maximum size: 125*140*205mm
Power supply: AC100-240V.50/60HZ
Fuse: 2 pieces of 220V0.5A, 2 pieces of 1.5A
Max power consumption: 25VA
Applicable infusion sets: Meet all kinds of standard brands of infusion sets.
Infusion rate range setting: 1-600ml/h
Electromagnetic wave protection: Meet the requirements of
IEC60601-1-2(2001 version)
Battery: Rechargeable lithium polymer battery 7.4V,
1600mA; working for 4 hours (10ml/h) after
fully being charged
Alarm function: over; OCCUL.; low battery; installation error; air; empty; No AC;power supply; abnormal control
Maximum weight: 2.8kg
Preset infusion range: 1-9999ml
Infusion increment: 1.0ml
Bolus infusion rate: 600ml/h
Accumulated infusion volume display: 1~9999ml
Infusion accuracy: ±3%( after being adjusted, up to ±2%)
KVO rate: 1ml/h
Operation condition: Environment temperature: 5℃~+40℃, air
pressure (86~106)kPa, relative humidity 80%
Storage condition: Environment temperature: -40℃~+55℃, air
pressure (50~106)kPa, relative humidity 95%
Panel instructions: infusion rate; infusion preset volume; accumulated infusion volume; battery
capacity; battery charging indication;AC power supply connection indication; over; OCCUL. ALARM; air; Bed No.