user login

Syringe pump/syringe driver, medical equipments

Min. Order: 1 Piece/Pieces
Trade Term: EXW
Payment Terms: T/T, WU
Supply Ability: 1000
Place of Origin: Guangdong

Company Profile

Location: Hong Kong
Business Type: Manufacturer, Distributor/Wholesaler

Product Detail

Model No.: Model BD-5000
Means of Transport: Ocean, Air
Production Capacity: 1000
Delivery Date: 7 working days
Show

Product Description

 Maximum injection rate:   1500ml/h (different syringe has different maximum flow rate)

Applicable syringe:       Automatic identification of 5ml, 10ml, 20ml, 30ml and 50ml.

Drug name;              such drugs as Penicillin and vancomycin can be shown; more drugs can be set according to the demand.

Power supply:             AC100-240V, 50/60HZ

Maximum weight:            <3.0kg

Battery:                Rechargeable lithium polymer battery 7.4V, 1600mA Working for 4 hours (25ml/h) after fully being charged

Max power consumption:    25VA

Accumulated volume display:0.1~999.9

Injection accuracy:            ±3%( after being adjusted, up to ±2%)

Rate increment:          0.1ml/h

Preset injection volume setting:0.1ml~9999.9ml

Preset time setting:        0-33h20 minutes

Classification:              class BF type

Injection rate range:         5ml: 0.1ml/h---100ml/h; 10ml: 0.1ml/h---200ml/h

                          20ml: 0.1ml/h---400ml/h; 30ml: 0.1ml/h---600ml/h

                           50ml: 0.1ml/h---1500ml/h

Bolus rate:             5ml:100ml/h10ml:200ml/h20ml:400ml/h30ml:400-600ml/h50ml:400-1500ml/h

Alarm:         over; injection near completion; no AC power supply; CPU communications error; pump stuck; disengaged.

Selectable modes:      1. Rate mode   2.time mode   3. Body weight mode

Time mode:                Volume limit: 0.1-999.9mltime: 1-2000min

Body weight mode:          Body weight: 0.1~300kg   drug limit: 0.1~999.9mg   volume limit: 0.1~999.9mlDose: 0.1~9999.9 (subject to unit, dosage, liquid, weight and syringe) unit: mg/kg/h, ug/kg/min

OCCUL value:               0.04Mpa-0.16Mpa, pressure change and the current phase against block valve position of syringe can be shown.

KVO rate:               Adjustable range: 0.1ml/h-5.0ml/h, choose "OFF" to cancel KVO function if no need KVO (entering into KVO after OCCUL when running)

Electromagnetic compatibility:Meet the requirements of IEC60601-1-2

(2001 version)

Electrical safety:             Meet the requirements of IEC60601-1:1995, GB9706.1-2007

Post Buying Request