7.Build-in communication interface to the central nurse station.
Specification
Fetal Heart Rate:
Transducer: Multi-crystals, pulsed doppler , high sensitively
Strength: <5mW/cm2
Working frequency: 1.0MHz
Signal processing: special DSP system and modern recognization.
Measurement range: 50~210 bpm
Alarm Range:
High limit: 160,170,180,190 bpm
Low limit: 90,100,110,120 bpm
Maximum audio output power : 1.5 Watt
Toco:
Measurement range: 0~100 units
Display
The LED displays FHR, TOCO,FM ,TIME
Power: a.c.100-240V,50/60Hz or 8.4 Li-LON
Power Consumption: < 20W
Dimension: 295W×240D×73H(mm)
Net weight:1.75kg