user login

Industrial Wastewater ORP Analyzer Probes, Oxidation Reduction Potential Meter PH Probes

Place of Origin: Zhejiang

Company Profile

Location: Beijing, China (Mainland)
Business Type: Manufacturer
Main Products: Waste Water Flow Meters

Product Detail

Model No.: ORP Analyzer

Product Description


Industrial Wastewater ORP Analyzer Probes, Oxidation Reduction Potential Meter PH Probes


1. Application

Monitoring of Oxidation-reduction potential in many fields, especially in municipal wastewater and industrial wastewater.

• water/wastewater treatment

Inflow of the plant, water collection well, aeration tank, the effluent, etc.

• pulp and paper, power plant, mine

water treatment, chemical treatment feeding

 

2. Features and Benefits

• The advanced large cross-section of Teflon diaphragm make the electrode bounds on liquid steadily.

• The unique structure makes the electrode have a long life duration and rarely clogged or polluted.

• The reference system filled with polymer, does not need electrolyte supplement.

• The electrode is applicable to all kinds of medium with serious contaminants, and is anti-interference to S2-, NH4+, CN-.

• History curve function: the measurement data is collected every 5 minutes, one month data can be stored.

• Notepad function, provides the record of operating and alarming time.

 

• When the electrode is calibrated, the mode, time and results are recorded, easy to review and analysis the change.

 

3. Specifications

Performance characteristics

Measuring Range

 -2000~+2000mv

Resolution

1mv

Accuracy

±0.05pH

Repeatability

±1%

Sensitivity

0.02pH

Calibration

Factory calibrated, also can be calibrated by customer on site

Display

LCD displays data including date, time, measurement value, trendline, etc.

Data storage

Automatic data store, power-off data protection

Outputs

Analog Outputs

0-10mA/4-20 mA output( maximum load 800 ohms)

Relay outputs

2 Relays with programmable hysteresis

Relay capacity

3A, 220VAC

Digital Communication

MODBUS RS485,Profibus DP is optional

Power supply

AC power supply

90~265 VAC, 50Hz

DC power supply

24 VDC ±10%

Physical parameters

Dimension of Transmitter

146×146×108mm(W×H×D)

(5.7*5.7*4.3 in.)

Dimension of Sensor

Φ28×165mm (1.1*6.5 in.)

Materials

--Transmitter: PC

--Sensor: Polyester

Post Buying Request