Place of Origin: | Zhejiang |
---|
Location: | Beijing, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Waste Water Flow Meters |
Model No.: | TCD10AC- TCD-S3C10 |
---|
Wastewater Digital Online Eectrical Conductivity Probe For Analyzer TCD10AC - TCD-S3C10
1. Application
Monitoring conductivity of surface water and wastewater, also could be used to for monitoring acid, alkali, salt concentration and total dissolved solids.
• Water/wastewater treatment
Inflow of the plant, water collection well, aeration tank, the effluent, etc.
• Pulp and paper, power plant, mine
Water treatment, chemical treatment feeding
2. Features and Benefits
• Large LCD screen with backlit.
• All kinds of data has password protection to avoid misoperation.
• History curve function: the measurement data is collected every 5 minutes, one month data can be stored.
• Notepad function, provides the record of operating and alarming time.
• When the electrode is calibrated, the mode, time and results are recorded, easy to review and analysis the change.
3. Specifications
Performance characteristics |
|
Measuring Range |
0~20000μs/cm |
Accuracy |
0.5%±1LSD |
Repeatability |
1% |
Sensitivity |
0.02pH |
Calibration |
Factory calibrated, also can be calibrated by customer on site |
Display |
LCD displays data including date, time, measurement value, trendline, etc. |
Data storage |
Automatic data store, power-off data protection |
Outputs |
|
Analog Outputs |
0-10mA/4-20 mA output( maximum load 750 ohms) |
Relay outputs |
2 Relays with programmable hysteresis |
Relay capacity |
3A, 220VAC |
Digital Communication |
MODBUS RS485,Profibus DP is optional |
Power supply |
|
AC power supply |
90~265 VAC, 50Hz |
DC power supply |
24 VDC ±10% |
Physical parameters |
|
Dimension of Transmitter |
146×146×108mm(W×H×D) (5.7*5.7*4.3 in.) |
Dimension of Sensor |
Φ28×98mm (1.1*3.9 in.) |
Materials |
--Transmitter: PC --Sensor: Polyester |
Process connection of sensor |
NPT 3/4 |
Weight |
--Transmitter: 1Kg --Sensor: 0.5Kg |
Length of s
|