Place of Origin: | Zhejiang |
---|
Location: | Beijing, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Waste Water Flow Meters |
Model No.: | MLSS10AC |
---|
0 - 9999mg/L 110V SS Analyzer, Suspended Solids Analyzer / Meter MLSS10AC
1. Application
Measure low, medium and high suspended solids concentrations using the multi-beam pulsed light method in municipal and industrial wastewater treatment process
• Sewage treatment plants
Inflow of wastewater, aeration tanks, returned sludge, primary sedimentation tanks, secondary sedimentation tank, thickening tank, sludge dewatering, etc.
• Water supply plants
Sedimentation tank
• Pulp & Paper
Pulp density
• Coal preparation plant
Sedimentation tanks, concentration tanks
• Power plant
Influent water treatment, Condensate, Boiler feed water, Cooling water, Wastewater
• Other industries:
Gush arenaceous liquid of PCB industry, marsh gas generate electricity( biogas power generation).
2. Features and Benefits
• Based on the innovative multi-beam light method for compensation of sensor soiling, the measurement is
not affected by contamination.
•High-quality 880nm optical and electronic assemblies eliminate interference from extraneous light sources.
• Digital sensor has good anti-interference performance and high stability, and the cable length is available in hundreds of meters.
• Connections between the sensor(s) and the transmitter are “plug and play”, and multi-channel transmitter operates either 1 or 2 sensors.
• Modbus RS485/RS232 or Profibus DP network connection for data transmission is Optional.
• The menu is available in Chinese or English.
• 3 Relays with programmable hysteresis provide high-limit alarm, low-limit alarm or timer.
• A built-in data logger with the capacity to store data for up to 10 years collects measurement data at user selectable intervals.
3. Specifications
Performance characteristics |
|
Measuring Range |
Activated sludge: 0~25g/L,0~9999mg/L, 0~400g/L (depends on the medium) |
Measurement Units |
g/L,,mg/L,ppm,% (optional) |
Resolution |
0.01g/L, 1mg/L, 1ppm, 0.01% |
Accuracy |
±1%FS (depends on the medium and homogeneity) |
Repeatability |
1% |
Response time |
Less than1 second (adjustable) |
Calibration |
Factory calibrated, also can be calibrated by customer on site |
Display |
LCD displays data including date, time, measurement value, trendline, etc., English menu |
Data storage |
Automatic Data Storage for up to 10 years, power-off data protection |
Outputs |
|
Analog Outputs |
0/4-20 mA output( maximum load 800 ohms), optional selection of 3.8mA or 21mA(under the condition of malfunction) |
Relay Outputs |
3 Relays with programmable hysteresis, alarm or timer function optional |
Relay Capacity |
2A, 250VAC |
Digi
|