Hardware requirements:
CPU:Pentium42.6G or above it
Internal memory: above 512MB
Hardware: above 80GB
Display card: 128MB Direct 3D compatible
Monitor : 17 inch pure level Liquid crystal display
Sound card: Direct sound compatible
Speaker:Multimedia active speaker
Printer:High performance ink jet printer or color printer
PCI serial card: The standard RS232interface
Network cable: Six core wire and nine DB head
Bedside monitor technical parameters:
Regular parameters:
Size: 300mm × 230mm × 130mm or 350mm × 240mm × 105mm
Weight: 3kg
Screen: 10.4inch TET colorful LCD screen or 12.1inch TET level colorful LCD screen
Resolution: 640 × 480 or 800 × 600
Standard of security:IEC601-1, UL2601
Running mode:running continuously
Power requirements:
The external AC power: 180~240V, the maximum current is 1 A Working temperature:0—40℃
Storting temperature: -20℃—50℃
Working humidity: the relative temperature is 5~95%,no Condensation
Storing humidity: the relative humidity is 5~95%,no condensation
Altitude: 0~100 foot
Measurement of Electrocardiogram:
Display: show the Seven channel ECG waveform at the same time
Mode: The filter bandwidth optional monitoring, diagnosis, operation
Gain option: four gears including ×1 \×2 \×4\×0.5
Lead-off detector: equipped
Calibration signal: 1mV
ST segment detection: range(-2.0mV----+2.0mV)
Anti interference: anti defibrillation, anti electric knife, anti interference electromyogram, anti interference of electrified wire netting
Range of heartbeat rate:25—250times/minute
Accuracy of heartbeat rate: ±2times/minute
Bandwidth:0.5HZ~100HZ
Lead: total lead, I, II, III, avF, avR, avL, V1-V6
Using standard: Conform to the standard of ANSI/AMI EC13-1992
Responding time:1s
Way of alarming: both in sound and light
Heart rate alarm preset range:upper limit,1~255times/minute
Lower limit,0~254times/minute
Breathing measurement:
Way of measuring: Impedance measurement principle
Measuring lead: RA-LA(I), A channel respiration wave
The range of the breathing rate:0~60times/minute
Precision; :±1 brpm
Responding time: 1s
Gain option: four gears including ×1\×2 \×4\×0.5
Alarming mode: both in sound and light
Alarm preset range:upper limit,10~100times/minute
Lower limit,0~99times/minute
Noninvasive blood pressure measurement:
It's principle: Oscillation method
Releasing mode: Ladder deflation
Security: conform to the requirement of CE
Measuring range:
**:Systolic pressure 30 ~ 255 mm hg
Average pressure: 20~ 235 mm hg
Diastolic pressure:15 ~ 220 mm hg
Pulse:30~240times/minute
New born baby: Systolic pressure 30 ~ 135 mm hg
Average pressure: 20 ~ 125 mm hg
Diastolic pressure: 15 ~ 110 mm hg
Pulse:40~240times/minute
Time spending: less than 45 seconds
Overvoltage limiting of software:
**:less than 290mmhg
New born baby:less than 145mmhg
Hardware protection:
**: 300 ± 10 mm hg
New born baby: 150 ± 4 mm hg
The initial inflation pressure value:
**:170 mm hg
New born baby: 65 mm hg
Sensor calibration:before measurement
Resolution: 1 mm hg
Precision: average deviation is less than ± 5 mm hg
The maximum average deviation is less than 8 mm hg
Pressure indicator accuracy: 3 mm hg
Leaking: less than 6 mm hg per minute
Subsequent cuff inflatable : Intelligent control
Measurement of temperature:
Technic: High sensitive thermistor probe
Range:20℃ — 45.5℃
Precision: ±0.1℃
Measurement: Single channel temperature measurement, the user can correct it freely
Responding time:1s
Way of alarming: sound and light
The range of alarm:
Upper limit: 20.1℃ - 45.5℃
Lower limit: 20℃ - 45.4℃
Measurement of oxygen saturation:
Range:0~100%
Precision:1%
Resolution:1%
Pulse: 20 — 250 times/minute
Responding time:1s
Way of alarming: sound and light
Range of alarm:
Upper limit: 1% - 100%
Lower limit: 0% - 99%
And central monitoring system also has the following functions:
The heartbeat of fetus
Probe: nine wafer, wide beam, the pulse working mode, high sensitivity
Ultrasonic intensity: less than 1 MW / cm2
Operating frequency:1.0MHz
Signal processing: Specialized digital signal processing unit
Maximum audio power output:1.5 W
Contraction stress:
Measuring range: 20- 100unit of pressure
Ultrasonic probe output: According to the international convention, the negative peak of sound pressure should be no more than 1 M Pa , acoustic beam of Sound intensity should be not more than 20 MW/cm2 ,Spatial peak of instantaneous average sound intensity density should be not more than 100 MW / cm 2. This type of sound intensity is not more than1mW/cm2
GEL:
Viscous water-based compound, without stimulating the skin, without allergy. It is stable chemically and a type of inhibiting bacteria.