Min. Order: | 1 Set/Sets |
---|---|
Payment Terms: | T/T |
Place of Origin: | Guangdong |
Location: | Shenzhen, Guangdong, China (Mainland) |
---|---|
Business Type: | Manufacturer |
Main Products: | Hematology Analyzer |
Model No.: | PE-6800 |
---|---|
Means of Transport: | Ocean, Air |
Brand Name: | prokan |
Technical Specifications: | ||
Principle: | Electrical resistance for counting, hemiglobincyanide method and SFT method for hemoglobin | |
Parameters: | 3-part differential of WBC; 20 parameters + 3 color histograms (WBC/RBC/PLT) | |
Sample Volume | Venous Mode: 9.6μL Capillary Mode: 9.6μLPre-diluted: 20μL | |
Working Mode: | Counting passages detection + independent HGB measurement system; 24hours with auto standby | |
Operation Mode: | Windows operation system with ,keyboard &mouse input | |
Throughput: | 35 samples/hour | |
Data Storage: | Up to 12,000 results inclusive of the histograms | |
QC mode: | L-J, ,X-B , SD, CV, QC histograms | |
Calibration Mode: | Relatively independent manual calibration and anto calibration in three detecting modes | |
Cross Contamination Ratio | WBC,RBC, HGB<0.5%, PLT<1% | |
Performance | Parameter Repeatability (CV %) | Linear Range |
WBC≤2.5% | 0.00-99.9×108/L | |
RBC ≤2.0% | 0.00-9.99×1012/L | |
HGB ≤1.5% | 0-300.0g/L | |
MCV≤0.4% | 40-150fL | |
PL T ≤5.0%(100-500) | 0-999×109/L | |
Display: | 8.4"LCD ,resolution:640*480 | |
Reference value | Baby,kid,women,man,normal and customization | |
Interface: | 2 × USB, 1 × parallel (for external printer), 1 × VGA (for external monitor), 2 × PS/2, 2 × RS-232(for network); | |
Printout | I Built-in thermal recorder, optional external printer, eight report formats | |
WorkingEnvironment | Temperature: 15℃~35℃; Humidity: 10%-90%; | |
Power Requirement | AC100-240V 50/60Hz |