user login

soft endoscope washer

Min. Order: 2 Set/Sets
Trade Term: FOB
Payment Terms: T/T
Supply Ability: 1000/year
Place of Origin: Guangdong

Company Profile

Location: China (Mainland)
Business Type: Manufacturer

Product Detail

Model No.: DE-660
Means of Transport: Ocean
Brand Name: MEIMEI
Production Capacity: 1000/year
Packing: wooden
Delivery Date: in 15days
Show

Product Description

Intelligent automatic design.
Wash and disinfect the whole surface of endoscope, use of water, medical enzymes, and disinfectant solution produced by high-pressure water.
Disinfectant solution can be Glutaraldehyde
Operation steps:
Wash, Rinse, Disinfect, Rinse again, Dry program.
DE-600 is a microcomputer-based fully automated disinfection system. All cleaning, disinfecting, drying and other procedures are standardized according to the rules and regulation of government health sector. Cleaning and disinfection are fully configurable with the accuracy up to seconds. This greatly ensures that cleaning and disinfection procedures are efficient and accurate. It not only avoids errors or insufficient disinfecting time introduced by manual operations, but also saves labor cost.
DE-600 is also proven to the best choice for endoscope disinfection. Natively by its fully automated disinfection process, health care personnel are well protected from the side effects of disinfectant since the direct and indirect contact with disinfectant is now completely avoidable
Parameters
DE-660
Water supply
press:0.2-0.4Mpa
Water drain
height≤350mm;pipe caliber≥50mm
Power supply
AC220V±10%,50Hz,max:900W
Disinfectant capacity
10L
Enzyme and alcohol
2.5L;2..5L
Leak testing
Leak testing during processing and alarm when leaking.
Disinfection record
Micro record after one disinfection
Disinfection time
glutaraldehyde solution:T≥41℃,10 min;T≥21℃,18 min;
1‰peroxyacetic :T≥41℃,8 min.
Alarm and
thermostatical set
Alarm when lack of disinfectant, machine out of order and display the fault; when under the demand temperature, the machine pauses
Electric lid
Avoid manual opening to cause infection
Noise level
≤50db(A)
Ambient temperature
0℃~40℃
Disinfection effect
None pathogenic bacteria,HbsAg negative,
average killing log value of germ>3.00
Self-disinfection
Avoid reinfection
Print
Size
W 580×D 610×H 915(mm)
Weight
58kg
Post Buying Request