Product Description
Description
DTJS-6009 is designed to measure the Capacitance and Dissipation Factor (tgδ) of dielectric and electric equipments. It can test grounded equipments, ungrounded equipments, and completely-sealed CVT (Capacitive Voltage Transformer).
The instrument generates testing High-voltage by itself, and the frequency is 45Hz and 55Hz to avoid electromagnetic interference from power frequency, so as to assure the measuring accuracy.
Features
‹ Non-power-frequency Testing voltage, Interference suppression circuit, to assure measuring Accuracy.
‹ Testing voltage is Program-controlled.
‹ Auto balance measuring bridge and calculate dissipation factor to speed and simplify test process.
‹ Menu operation.
‹ Built-in Printer.
‹ Auto identify earthing to ensure safety.