Ventilator
■Display mode
High-definition 5.7" LCD screen display
■Ventilation mode
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
■Ventilation parameters
Tidal volume: 50~1500ml
Rate: 2~99bpm
SIMV rate: 2~20bpm
I:E: 2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
PEEP: 0~2.0 kPa
Pressure range: 1.0~6.0 kPa
SIGH: 1.5 times the inspiratory time
■ Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure
■Security Alarm System
Airway pressure alarm
upper limit setting range: 1.0~6.0 kPa
low limit setting range: 0.4~2.0 kPa
Per-minute ventilation volume alarm
upper limit setting range: 3.0~30L/min
low limit setting range: 1.0~10L/min
Sustained high-pressure alarm
it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm
it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
■ Power
AC 220V±10% 50Hz±1Hz
UPS, Storage battery