user login

Syringe pump

500.0 USD
Min. Order: 1 Set/Sets
Payment Terms: T/T
Supply Ability: 6000 sets/year
Place of Origin: Guangdong

Company Profile

Location: China (Mainland)
Business Type: Manufacturer

Product Detail

Model No.: HX-901A
Means of Transport: Air or Sea
Production Capacity: 6000 sets/year
Packing: Corrugated carton
Delivery Date: 10 days
Show

Product Description

1.Syringe Mode
Micro and continuous injection
2.Syringes
10mL, 20mL, 30mL, 50/60mL syringes, self-identified automatically.
3.Syringe Brands
10 usual syringe brands preset, and easily customize for others.
4.Flow Rate Range
10mL syringe: 0.1 ~ 300 mL/h
20mL syringe: 0.1 ~ 400 mL/h
30mL syringe: 0.1 ~ 600 mL/h
50mL syringe: 0.1 ~ 1200 mL/h
5.Increment
0.1mL/h
6.Flow Rate Accuracy
≤ ± 2% (including accuracy ± 1% on mechanism)
7.Volume Limit
0.1~ 9999.9 mL and empty mode available
8.Volume Infused Display
0.1 ~ 9999.9 mL
9.Purge Rate
10mL syringe: 300 mL/h
20mL syringe: 400 mL/h
30mL syringe: 600 mL/h
50mL syringe: 1200 mL/h
10.Bolus rate
10mL syringe: 0.1 ~ 300 mL/h
20mL syringe: 0.1 ~ 400 mL/h
30mL syringe: 0.1 ~ 600 mL/h
50mL syringe: 0.1 ~ 1200 mL/h
11.Optional Inject Rate Unit
mL/h, mg/h, µg/kg/min, mg/kg/min, µg/kg/h, mg/kg/h, and µg/h.
12.KVO Rate
1 mL/h
13.Occlusion Pressure
Pressure detection device with 8 sensitivities setting, 20 ~ 140KPa, dynamically display pressure value.
14.Audio and Visual Alarms
AC power disconnected, Syringe barrel clasp, Plunger head position, Injection finished, End of injection pre-alarm, Empty syringe, Occlusion, Low battery, Battery run out, Battery damaged, No operation, System failed.
15.Display
LCD with blue backlight
16.Power Source
AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA
17.Battery
Rechargeable Ni-MH battery: DC9.6V, 2000mAh, life: ≥ 4 hours at 5mL/h, remaining battery displayed.
18.Classification
 CF, IPX4
19.Dimensions and N.W.
335mm x 220mm x 205mm; 2.0kg
Post Buying Request